Skip to content

CARBOMEDICS CARBO-SEAL AP-027Recalls & reported problems

Corcym Srl · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08022057013491), received between Jun 28, 2026 and Jun 28, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 28, 2026Insufficient Device Problem Information

    THE MANUFACTURER WAS NOTIFIED OF AN INTRAOPERATIVE EXPLANT INVOLVING CARBOSEAL ASCENDING AORTIC PROSTHESIS THROUGH DEVICE TRACKING. BASED ON THE INFORMATION AVAILABLE, A CARBOSEAL ASCENDING AORTIC PROSTHESIS AP-027 WAS ATTEMPTED TO BE IMPLANTED AS FOLLOWS: THE SURGEON WAS DOING A BENTALL PROCEDURE (VALVE/AORTIC ROOT REPLACEMENT). SIZED FOR A CARBOSEAL 27 (AP-027). PUT THE SUTURES IN THE ANNULUS IN THE NORMAL FASHION, FOLLOWED BY PUTTING THEM THROUGH THE SEWING CUFF OF THE VALVE. LOWERED IT DOWN INTO PLACE AND STARTED TYING IT IN ON ONE SIDE. WHEN THE SURGEON BEGAN TYING THE VALVE IN ON THE OPPOSITE SIDE, A POP WAS FELT. THROUGH INSPECTION, IT WAS NOTICED THAT THE LVOT ON THE LEFT SIZE HAD BEEN TORN AND THERE WAS THEN A HOLE. AS SUCH, THEY HAD TO REMOVE THE CARBOSEAL, REPAIR THE TISSUE AND AT THAT TIME DECIDED TO IMPLANT A GRAFT (TERUMO GELWEAVE), FOLLOWED BY A TOP HAT AORTIC VALVE S5-025 INSIDE THE GRAFT TO KEEP IT SUPRA-ANNULAR AND AWAY FROM THE REPAIR ON THE HEART AS WELL AS THE GRAFT ANASTOMOSIS. THE PATIENT RECOVERED FINE AND HAS SINCE BEEN DISCHARGED HOME. BASED ON THE MEDICAL JUDGMENT RECEIVED, THE PATIENT HAD VERY POOR-QUALITY TISSUES WHICH WERE RELATED TO THE TEAR.

    Patient problems: Insufficient Information

    MAUDE report 25647273 ↗

Other HEART-VALVE, MECHANICAL devices

All HEART-VALVE, MECHANICAL devices →

Sources and updates