Paradym RF DR 9550Recalls & reported problems
Microport Crm Srl · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 08031527012254), received between Apr 25, 2017 and Apr 30, 2021 — classified as 9 injuries.
The most frequently reported specific device problem is “Contamination /Decontamination Problem” (2 reports).
4 other recalls cover devices cleared under the same FDA submission (P980049); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Contamination /Decontamination Problem2
- Over-Sensing2
- Inappropriate or Unexpected Reset1
- Inappropriate/Inadequate Shock/Stimulation1
- Low impedance1
- Premature Elective Replacement Indicator1
Patient problems
- Unspecified Infection4
- No Information3
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryApr 30, 2021Premature Elective Replacement Indicator; Premature Elective Replacement Indicator
REPORTEDLY, THE RECOMMENDED REPLACEMENT TIME (RRT) WAS REACHED EARLIER THAN EXPECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11753704 ↗InjuryFeb 8, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REPORTEDLY, AN INFECTION RELATIVE TO THE SUBJECT DEVICE OCCURRED. THE DEVICE WAS EXPLANTED ON (B)(6) 2020.
Patient problems: Unspecified Infection
MAUDE report 11295696 ↗InjuryOct 20, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REPORTEDLY, ON (B)(6) 2020, THE DEFIBRILLATION SYSTEM WAS EXPLANTED DUE TO AN INFECTION. PRELIMINARY ANALYSIS SHOWED THAT THE SUBJECT DEVICE WAS STERILIZED AND RELEASED ACCORDING TO ALL APPLICABLE STANDARD OPERATING PROCEDURES.
Patient problems: Unspecified Infection
MAUDE report 10707549 ↗InjuryJul 20, 2020Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Low impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Low impedance
REPORTEDLY, THE PATIENT WAS ADMITTED TO THE HOSPITAL AFTER HAVING RECEIVED SEVERAL INAPPROPRIATE SHOCKS BOTH AT HOME AND AT THE HOSPITAL. A MAGNET WAS APPLIED ON THE ICD UNTIL A COMPANY REPRESENTATIVE COULD INTERROGATE THE ICD. DEVICE INTERROGATION REVEALED THAT VENTRICULAR COIL CONTINUITY MEASUREMENTS WERE BELOW 200 OHMS. SINCE (B)(6) 2020, NON-PHYSIOLOGICAL SIGNALS WITH FAST RHYTHM WERE OBSERVED ON THE VENTRICULAR CHANNEL, LEADING TO THE DELIVERY OF INAPPROPRIATE THERAPIES SINCE (B)(6) 2020. THOUGH THE RIGHT LEAD IMPEDANCE CURVE WAS RELATIVELY STABLE, PUNCTUAL VARIATIONS OF MORE THAN 200 OHMS WERE NOTED SINCE (B)(6) 2020. CONSEQUENTLY, IT WAS DECIDED TO TURN OFF THE THERAPIES. A REVISION OF THE DEFIBRILLATION SYSTEM AND ITS REPLACEMENT WERE PERFORMED ON (B)(6) 2020, AS THE CONSECUTIVE SHOCK DELIVERIES COULD HAVE INDUCED A DECREASE OF THE BATTERY LONGEVITY. PRELIMINARY ANALYSIS REVEALED THAT THE INAPPROPRIATE SHOCKS DELIVERED ON (B)(6) 2020 MOST PROBABLY RESULTED FROM A VENTRICULAR LEAD ISSUE.
Patient problems: No Information
MAUDE report 10295421 ↗InjuryOct 2, 2018Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset
REPORTEDLY, UPON INTERROGATION OF THE DEVICE, MULTIPLE WARNING MESSAGES WERE DISPLAYED INCLUDING: [27] THE DEVICE WAS REINITIALIZED 1 TIME SINCE THE BEGINNING OF LIFE, [59] RESET OCCURRED ON (B)(6) 2018, [A1] LOW SHOCK IMPEDANCE ON (B)(6) 2018 AND [A4] MAX SHOCK ENERGY INEFFECTIVE ON (B)(6) 2018. THE PATIENT IS HOSPITALIZED AND REPORTEDLY MAY HAVE BEEN EXPERIENCING RUNS OF VT AND MAY HAVE BEEN RECEIVING SHOCKS. IT IS ALSO POSSIBLE THE PATIENT WAS EXTERNALLY CARDIOVERTED HOWEVER, THIS WAS NOT CONFIMED. PRELIMINARY ANALYSIS REVEALED THAT OVERLOAD CONDITIONS WERE MET ON (B)(6) 2018. AND CONFIRMED THAT THE RESET OCCURRED. THIS RESET WAS RELATED TO THE DETECTION OF CORRUPTED DATA IN THE DEVICE MEMORIES.
Patient problems: No Information
MAUDE report 7925078 ↗InjuryDec 4, 2017Over-Sensing; Over-Sensing
REPORTEDLY, DURING A FOLLOW-UP PERFORMED (B)(6) 2017, EPISODES STORED IN THE DEVICE MEMORIES WERE SUSPICIOUS OF VENTRICULAR OVERSENSING.
Patient problems: No Information
MAUDE report 7079312 ↗InjuryJun 1, 2017Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT THE SUBJECT DEVICE AND THE ASSOCIATED LEADS WERE REMOVED DUE TO BLOOD INFECTION.
Patient problems: Unspecified Infection
MAUDE report 6605318 ↗InjuryMay 17, 2017Contamination /Decontamination Problem; Contamination /Decontamination Problem
REPORTEDLY, THE DEFIBRILLATION SYSTEM WAS EXPLANTED DUE TO INFECTION. THE SUBJECT DEVICE WILL BE RETURNED FOR ANALYSIS. AN INVESTIGATION IS REQUIRED. PRELIMINARY ANALYSIS SHOWED THAT THE DEVICE HAS BEEN STERILIZED AND RELEASED ACCORDING TO ALL APPLICABLE PROCEDURES.
Patient problems: Unspecified Infection
MAUDE report 6571898 ↗InjuryApr 25, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REPORTEDLY, THE SUBJECT ICD WAS EXPLANTED WITH LEADS DUE TO DUE TO COMPLICATIONS. IT WAS MENTIONED ALSO 'DEFECTIVE LEAD'. THE SUBJECT ICD WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6519045 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P980049. One submission often covers many models, so check the product and lot details in each record.
- Z-3231-2017 · Jul 24, 2017 · TerminatedPlatinium VR DF4 1240, Model Number TDF035U, UDI/GTIN 08031527015460, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single
- Z-3232-2017 · Jul 24, 2017 · TerminatedPlatinium DR DF4 1540, Model Number TDF036U, UDI/GTIN 08031527015477, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single
- Z-3229-2017 · Jul 24, 2017 · TerminatedPlatinium DR 1510, Model Number TDF032U, UDI/GTIN 08031527015521, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single cham
- Z-3228-2017 · Jul 24, 2017 · TerminatedPlatinium VR 1210, Model Number TDF031U, UDI/GTIN 08031527015514, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single cham
Other DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER devices
- Rivacor 7 DR-T DF4 ProMRI 429534BIOTRONIK SE & Co. KG · DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER2 deaths286 reports
- Iperia 7 DR-T DF4 ProMRI 392423BIOTRONIK SE & Co. KG · DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER3 deaths256 reports
- Ilivia 7 DR-T DF4 ProMRI 404623BIOTRONIK SE & Co. KG · DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER2 deaths233 reports
- Iperia 7 DR-T DF-1 ProMRI 392409BIOTRONIK SE & Co. KG · DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER93 reports
- Ilesto 7 DR-T DF-1 US 383562BIOTRONIK SE & Co. KG · DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER79 reports
- Ilivia 7 DR-T DF-1 ProMRI 404622BIOTRONIK SE & Co. KG · DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER74 reports
- Ilivia Neo 7 DR-T DF-1 ProMRI 429529BIOTRONIK SE & Co. KG · DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER1 death56 reports
- Itrevia 7 DR-T DF4 392426BIOTRONIK SE & Co. KG · DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER1 death48 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page