Skip to content

BlueSky TPM020URecalls & reported problems

Microport Crm Srl · BlueSky DR Alizea 1600 · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08031527018997), received between Jan 17, 2025 and Jan 17, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Unexpected Mode Switch” (2 reports).

1 other recall cover devices cleared under the same FDA submission (P950029); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unexpected Mode Switch2
  • Premature Discharge of Battery1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 17, 2025Premature Discharge of Battery; Unexpected Mode Switch; Premature Discharge of Battery; Unexpected Mode Switch

    ON (B)(6), DURING AN IN-CLINIC INTERROGATION OF ALIZEA DR (B)(6), A WARNING MESSAGE STATING "RRT OCCURRED ON (B)(6) 2024. MAGNET RATE 80" IS DISPLAYED. THE DEVICE IS FOUND PROGRAMMED IN VVI MODE AND IS DISPLAYING THE CURRENT BATTERY VOLTAGE VALUE TO BE 3.12V.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21179847 ↗
  • InjuryJan 17, 2025Unexpected Mode Switch; Unexpected Mode Switch

    ON DECEMBER 19, 2024, MICROPORT CRM REP (B)(6) CALLED TECHNICAL SERVICES STATING THAT HE WAS PERFORMING AN IN-CLINIC INTERROGATION OF (B)(6). UPON INTERROGATION, A WARNING MESSAGE STATING "RRT OCCURRED ON (B)(6) 2024. MAGNET RATE 80" IS DISPLAYED. (B)(6) STATES THE DEVICE IS FOUND PROGRAMMED IN VVI MODE AND IS DISPLAYING THE CURRENT BATTERY VOLTAGE VALUE TO BE 3.12V.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21179596 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P950029. One submission often covers many models, so check the product and lot details in each record.

  • Z-2235-2009 · Sep 23, 2008 · Terminated
    Sorin Group Reply DR Rate responsive dual-chamber pacemaker. Made in Italy by: Sorin BioMedica CRM S.r.l., Via Crescentino s.n., 13040 Saluggia (VC) Italy. ELA Medical, Inc., 14

Other Implantable Pacemaker Pulse-Generator devices

All Implantable Pacemaker Pulse-Generator devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page