CARPEDIEM IB0595570Recalls & reported problems
Bellco Srl · BL 250 HEMOFILTRATION/HEMODIALYSIS KIT 015 FOR CARPEDIEM · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08032979792282), received between Jan 5, 2022 and Jan 5, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak1
- Leak/Splash1
- Protective Measures Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 5, 2022Fluid/Blood Leak; Leak/Splash; Protective Measures Problem; Fluid/Blood Leak; Leak/Splash; Protective Measures Problem
ACCORDING TO THE REPORTER, THE MACHINE WAS SET UP FOR A PATIENT TREATMENT AND DURING THE PRIMING PHASE THE NURSE NOTICED THAT THE FLUID WAS LEAKING FROM THE PRE-FILTER PRESSURE TRANSDUCER AREA. PRIMING WAS STOPPED AND THE PREASSEMBLED DEVICE WAS TAKEN DOWN, AND A SECOND DEVICE FROM THE SAME LOT WAS USED AND HAD THE SAME ISSUE AND SO IT WAS REPLACED WITH ANOTHER PRODUCT. THEN A NEW TREATMENT WAS STARTED WITH THE NEW PREASSEMBLED DEVICE., IT WAS ALSO STATED THAT DURING NORMAL PATIENT THERAPY, USER NOTICED A LARGE BLOOD LEAK (WITHIN THE PREASSEMBLED DEVICE, WITHIN THE DIALYZER AND ALSO IN THE EFFLUENT LINE AND INF/DIA LINE CLOSEST THE DIALYZE) . THE THERAPY WAS STOPPED EARLY, AND THE PATIENT WAS DISCONNECTED FROM THE MACHINE, THE PREASSEMBLED DEVICE WAS ALSO TAKEN DOWN, AND A NEW TREATMENT WAS STARTED WITH A NEW PREASSEMBLED DEVICE TO RESOLVE THE ISSUE AND THE TREATMENT WAS COMPLETED. ALARMS WERE ACTIVATED (FIRST OCCURRENCE DURING PRIMING- ALARM 33 MINIMUM PRE-FILTER PRESSURE WAS DISPLAYED. DURING THERAPY ALARM 28 BLOOD LEAKAGE FROM FILTER WAS DISPLAYED.). NO OTHER DEFECTS/DAMAGES FOUND ON THE PRODUCT WHERE THE LEAK WAS OBSERVED, INTERVENTION/TREATMENT WAS NOT REQUIRED AS A RESULT OF
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13172183 ↗
Other Pediatric Hemodialysis System devices
- CARPEDIEM IB7010250Bellco Srl · Pediatric Hemodialysis System3 deaths13 reports
- CARPEDIEM IB0595580Bellco Srl · Pediatric Hemodialysis System1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗