Skip to content

SMR Shoulder Humeral Head (REF 1322.09.480)Recalls & reported problems

Limacorporate SPA · Humeral heads (CoCrMo) dia.48mm · Class 2 · Implant

All models of SMR Shoulder (67)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033390001403), received between Jan 22, 2026 and Jan 22, 2026 — classified as 1 injury.

2 other recalls cover devices cleared under the same FDA submission (K100858); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Implant Pain1
  • Muscle/Tendon Damage1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 22, 2026Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY PERFORMED ON (B)(6) 2025. PATIENT PRESENTED SEVERAL YEARS POST-OP WITH COMPLAINTS OF PAIN. SURGEON CONVERTED THE PROSTHETIC ASSEMBLY FROM AN ANATOMIC CONFIGURATION TO A REVERSE DUE TO NO LONGER INTACT ROTATOR CUFF. SURGEON REMOVED THE PRE-EXISTING HUMERAL HEAD (PART NUMBER 1322.09.480), NEUTRAL ADAPTOR TAPED (1330.15.270), FINNED HUMERAL BODY (PART NUMBER 1350.15.110) AND POLYETHYLENE LINER FOR METAL BACK (PART NUMBER 1377.50.005), AND IMPLANTED THE REVERSE CONFIGURATION CONSISTING OF A 36MM ECCENTRICAL GLENOSPHERE (PART NUMBER 1376.09.031) WITH LATERALIZED CONNECTOR (PART NUMBER 1374.15.314), A REVERSE 140° HUMERAL BODY (PART NUMBER 1352.15.011) AND REVERSE LINER +3MM (PART NUMBER 1360.50.815). SURGERY WAS COMPLETED AS INTENDED. PREVIOUS SURGERY IS UNKNOWN, AS WELL AS LOT NUMBER OF INVOLVED COMPONENTS. PATIENT IS MALE, DATE OF BIRTH FEBRUARY 22ND, 1950. THE EVENT OCCURRED IN THE UNITED STATES.

    Patient problems: Implant Pain; Muscle/Tendon Damage

    MAUDE report 24144121 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K100858. One submission often covers many models, so check the product and lot details in each record.

  • Z-2540-2015 · Aug 14, 2015 · Terminated
    SMR Shoulder/Spalla Trial Heads ECC .4 mm Adaptor, REF 9013.30.021; SMR Shoulder/Spalla Trial Heads ECC .8 mm Adaptor, REF 9013.30.031; for use in shoulder implant surgery.
  • Z-2539-2015 · Aug 14, 2015 · Terminated
    SMR Shoulder/Spalla Trial Humeral Head diameter .42 mm, REF 9013.22.425; SMR Shoulder/Spalla Trial Humeral Head diameter .44 mm, REF 9013.22.445; SMR Shoulder/Spalla Trial Humeral

Other Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented devices

All Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page