Skip to content

Delta Dual Mobility System 5885.09.042Recalls & reported problems

Limacorporate SPA · Liner #L for Mobile Liner Dia. 42 mm · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08033390020060), received between Jul 18, 2024 and Feb 25, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Mechanical Problem1

Patient problems

  • Implant Pain1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 25, 2025Mechanical Problem; Mechanical Problem

    DURING HIP SURGERY PERFORMED ON (B)(6) 2025, AN ISSUE WAS OBSERVED INTRAOPERATIVELY WITH THE LINER #L FOR MOBILE LINER Ø42 (PRODUCT CODE: 5885.09.042, BATCH N° 2430719, STERILIZATION N° 2400269) DESPITE BEING CORRECTLY POSITIONED, THE INSERT WAS MOBILIZING. ADDITIONALLY, THE COMPONENT, ALTHOUGH INITIALLY APPEARING STABLE, WOULD DISASSEMBLE FROM THE CUP WHEN MANEUVERS WERE PERFORMED. DUE TO THESE ISSUES, IT WAS NECESSARY TO USE ANOTHER POLYETHYLENE INSERT. PATIENT: MALE EVENT OCCURRED IN ITALY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21459560 ↗
  • InjuryJul 18, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    HIP REVISION SURGERY DUE TO LOOSENING OF THE LINER #L FOR MOBILE LINER Ø42 (PRODUCT CODE 5885.09.042, LOT 2115888, STER. (B)(4)) AND DELTA NEUTRAL SPACER #XL (PRODUCT CODE 5885.15.610, LOT 2014107, STER. (B)(4)). THE DEVICES WERE USED IN COMBINATION WITH THE FOLLOWING DEVICES: - DELTA MULTIHOLE TT CUP Ø74MM - LINER #XL - TI6AL4V - REF. (B)(4) - LOT 2005209, STER. (B)(4) - BONE SCREW Ø6,5 H.20MM - TI6AL4V - REF. (B)(4) - LOT 2200053, STER. (B)(4) - BONE SCREW Ø6,5 H.30MM - TI6AL4V - REF. (B)(4) - LOT 2126547, STER. (B)(4) - BONE SCREW Ø6,5 H.35MM - TI6AL4V - REF. (B)(4) - LOT 2124322, STER. (B)(4) - MOBILE LINER ØINT 28 MM Ø42 MM - UHMWPE - REF. (B)(4) - LOT 21AT08K, STER. (B)(4) - HEAD FROM ANOTHER MANUFACTURER - STEM FROM ANOTHER MANUFACTURER THE REVISION SURGERY WAS PERFORMED ON (B)(6) 2024, THE PREVIOUS SURGERY TOOK PLACE ON (B)(6) 2022. ACCORDING TO THE AVAILABLE INFORMATION PROVIDED BY THE COMPLAINT SOURCE, THE PATIENT HAD A POSITIVE EVOLUTION AND RECOVERY AFTER THE FIRST SURGERY, HE FOLLOWED POST-OPERATIVE RECOMMENDATION AND WAS LEADING A NORMAL LIFE. IT WAS ALSO REPORTED THAT THE PATIENT HAD ALWAYS HEARD "CLICKS" WHEN SITTING DOWN BUT WAS TOLD THIS WAS NORMAL. ON (B)(6) 2024

    Patient problems: Implant Pain

    MAUDE report 19772699 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices →