Skip to content

SMR System 1362.09.120Recalls & reported problems

Limacorporate SPA · SMR System - Reverse HP Lateralizing Liner Long Dia. 44mm · Class 2 · Implant

All models of SMR System (47)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033390050036), received between Sep 23, 2026 and Sep 23, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Unstable” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unstable1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 23, 2026Unstable

    A revision surgery was performed on (b)(6) 2025 due to implant instability. The surgical approach was restricted in terms of lateralisation on the glenoid side due to the monoblock design of the custom glenoid. Different humeral options were trialled to achieve stability. The surgeon felt that the extension and long lateralised liner were providing excessive length without corresponding lateralisation, potentially contributing to instability. The extension and long lateralised liner were therefore removed and replaced with a long lateralised liner.Previous surgery was performed on (b)(6) 2025. The CMD 25-1154 implant (CMD 25-1154 Monoblock Glenoid, product code 9617.PE.04Z), which had been implanted on (b)(6) 2025, was not revised, together with the humeral body and the stem.The surgeon followed the standard surgical workflow for shoulder revision. The explanted components included:SMR HUMERAL EXTENSION + 9 MM (Product code 1352.15.001, lot 2422607, ster. n. 2400222).SMR REVERSE HP LAT. LINER LONG (Product code 1362.09.120, lot 2422590, ster. n. 2400215).Patient is Male, 81 years old.Event happened in Australia.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50865350 ↗

Other Shoulder Prosthesis, Reverse Configuration devices

All Shoulder Prosthesis, Reverse Configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page