The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033390094856), received between Feb 18, 2026 and Feb 18, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Mechanical Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 18, 2026Mechanical Problem

    DURING A SHOULDER PROSTHESIS REVISION PERFORMED ON (B)(6) 2026, AN INTRA-OPERATIVE ISSUE RELATED TO THE INSTRUMENTATION WAS ENCOUNTERED. THE SURGEON EXPERIENCED DIFFICULTIES USING TWO BASEPLATE EXTRACTOR SMALL-R DEVICES (PRODUCT CODE 9013.75.391; LOT 21BH07Z AND LOT 23BH03A) TO REMOVE THE BASEPLATE FROM THE TT PEG. SPECIFICALLY, THE SPLINES OF THE EXTRACTORS BENT DURING THE EXTRACTION ATTEMPTS, MAKING IT IMPOSSIBLE TO DISENGAGE THE BASEPLATE FROM THE PEG. IN ADDITION, THE SMALL-R CANNULATED REAMER (PRODUCT CODE 9013.75.395; LOT 15AQ00R), INTENDED FOR TT PEG REMOVAL, WAS DIFFICULT TO USE, PARTLY DUE TO THE ABSENCE OF AN APPROPRIATE LEVERAGE HANDLE. AS A RESULT OF THESE DIFFICULTIES, AN ALTERNATIVE INSTRUMENT (BRISTOW) WAS USED; HOWEVER, DESPITE MULTIPLE ATTEMPTS, THE PEG COULD NOT BE REMOVED. THE SURGEON SUBSEQUENTLY MODIFIED THE SURGICAL PLAN: THE ORIGINAL PEG WAS LEFT IN SITU, AND A NEW BASEPLATE WITH NEW SCREWS WAS IMPLANTED ONTO THE EXISTING PEG. THE SIZE OF THE GLENOSPHERE WAS ALSO REDUCED COMPARED TO THE INITIAL PLAN. THE OVERALL PROCEDURE TIME WAS PROLONGED BY APPROXIMATELY 45 MINUTES. THE EVENT OCCURRED IN AUSTRALIA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24377809 ↗

Other Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented devices