Skip to content

SMR Shoulder Expansion ExtractorRecalls & reported problems

Limacorporate SPA · Expansion Extractor · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08033390097598), received between Jan 7, 2025 and Oct 23, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

2 other recalls cover devices cleared under the same FDA submission (K100858); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 23, 2025Break; Break

    DURING A SHOULDER REVISION SURGERY PERFORMED ON (B)(6) 2025, WHEN USING THE SMR - EXPANSION EXTRACTOR (PRODUCT CODE 9013.52.165, LOT #15AA052). WHEN USING THE ER REMOVAL TOOL FOR THE HUMERAL ADAPTOR, THE PEGS WERE NOT STRAIGHT. IT ENDED UP BREAKING OFF AT THE END. SURGICAL TIME EXTENDED BY 20 MINUTES. THE SURGICAL TIME WAS EXTENDED BY 20 MINUTES DUE TO THE ISSUE. EVENT HAPPENED IN AUSTRALIA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23365752 ↗
  • MalfunctionFeb 5, 2025Break; Break

    DURING A SHOULDER REVISION SURGERY PERFORMED ON (B)(6) 2025, WHEN USING THE SMR - EXPANSION EXTRACTOR (PRODUCT CODE 9013.52.165, LOT #15AA052) IN THE HUMERAL BODY, TWO OF THE EXPANSION TINES BROKE. IT WAS REPORTED THAT ONE WAS IMMEDIATELY RECOVERED, THE OTHER WAS NOT NOTICED UNTIL TRYING TO SEAT THE HUMERAL REAMER GUIDE. IT WAS CONFIRMED THAT NO RESIDUAL WAS LEFT IN THE PATIENT. THE STEM EXTRACTOR (PRODUCT CODE 9013.02.301) WAS USED TO COMPLETE THE SURGERY. THE SURGICAL TIME WAS EXTENDED BY 15 MINUTES DUE TO THE ISSUE. THE INSTRUMENT HAS BEEN USED MORE THAN 15 TIMES. EVENT HAPPENED IN AUSTRALIA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21311499 ↗
  • MalfunctionJan 7, 2025Break; Break

    DURING A SHOULDER SURGERY ON (B)(6) 2024, TWO OUT OF THE 4 PRONGS OF THE FOLLOWING INSTRUMENT BROKE OFF: - SMR - EXPANSION EXTRACTOR (PRODUCT CODE 9013.52.165, LOT NUMBER 14AA253) THE SURGERY WAS COMPLETED REMOVING THE ENTIRE STEM CONNECTED TO PROXIMAL BODY. THE INSTRUMENT HAS BEEN USED 5 TIMES APPROXIMATELY. THERE WAS NO EXTENSION OF THE SURGICAL TIME. THE PATIENT IS A MALE, IN ITS 70S. THE EVENT HAPPENED IN THE UNITED STATES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21091029 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K100858. One submission often covers many models, so check the product and lot details in each record.

  • Z-2540-2015 · Aug 14, 2015 · Terminated
    SMR Shoulder/Spalla Trial Heads ECC .4 mm Adaptor, REF 9013.30.021; SMR Shoulder/Spalla Trial Heads ECC .8 mm Adaptor, REF 9013.30.031; for use in shoulder implant surgery.
  • Z-2539-2015 · Aug 14, 2015 · Terminated
    SMR Shoulder/Spalla Trial Humeral Head diameter .42 mm, REF 9013.22.425; SMR Shoulder/Spalla Trial Humeral Head diameter .44 mm, REF 9013.22.445; SMR Shoulder/Spalla Trial Humeral

Other Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented devices

All Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented devices →