Skip to content

PHYSICA Instrument Set Fem. Positioner Inserter/ExtractorRecalls & reported problems

Limacorporate SPA · Femoral Positioner Inserter/Extractor · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033390097956), received between Feb 13, 2025 and Feb 13, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 13, 2025Break; Break

    DURING A KNEE SURGERY PERFORMED ON (B)(6) 2025, THE SCREW FROM THE FEMORAL POSITIONER FELL OUT DURING IMPLANTATION OF FEMUR. THE INSTRUMENT INVOLVED IN THE EVENT IS THE FOLLOWING: PHYSICA - FEMORAL POSITIONER (PART CODE 9065.88.120, LOT NUMBER 17AG083). THE SCREW WAS NOTICED ON THE POST-OP X-RAYS. THEN THE SURGEON DECIDED TO TAKE THE PATIENT BACK TO OR TO RETRIEVE THE SCREW. THE PATIENT IS A MALE, DATE OF BIRTH (B)(6) 1952. THE EVENT HAPPENED IN THE UNITED STATES.

    Patient problems: Foreign Body In Patient

    MAUDE report 21374165 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page