SMR Shoulder Liner (REF 1365.50.815)Recalls & reported problems
Limacorporate SPA · Reverse Liner 40mm +3mm · Class 2 · Implant
All models of SMR Shoulder (65)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 08033390102933), received between Apr 4, 2025 and Jul 2, 2026 — classified as 8 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated7
- Adverse Event Without Identified Device or Use Problem1
- Patient Device Interaction Problem1
- Unstable1
Patient problems
- Joint Dislocation7
- Bone Fracture(s)1
- Joint Laxity1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryJul 2, 2026Device Dislodged or Dislocated
REVISION SURGERY OCCURRED ON (B)(6) 2026, DUE TO SHOULDER DISLOCATION. PRE-EXISTING FOLLOWING COMPONENTS HAS BEEN EXPLANTED: SMR REVERSE LINER +3MM D.40MM (PART NUMBER 1365.50.815, LOT. 23AT24A, STER. 2400018); SMR GLENOSPHERE Ø 40MM (PART NUMBER 1374.09.121, LOT. 2415894, STER. 2400152); SMR CONNECTOR SMALL R (PART NUMBER 1374.15.305, LOT. 2325339, STER. 2300250); AND REPLACED WITH NEW FOLLOWING ONES: SMR REVERSE LINER RETENTIVE +3MM D.40MM (PART NUMBER 1365.50.816); SMR GLENOSPHERE Ø 40MM (PART NUMBER 1374.09.121); SMR CONNECTOR LAT. +4MM SMALL R (PART NUMBER 1374.15.314). PREVIOUS SURGERY HAS BEEN PERFORMED ON (B)(6) 2024. PATIENT IS MALE, DATE OF BIRTH (B)(6) 1943. EVENT OCCURRED IN THE UNITED STATES.
Patient problems: Joint Dislocation
MAUDE report 25687433 ↗InjuryJan 26, 2026Device Dislodged or Dislocated
REVISION SURGERY WAS PERFORMED ON (B)(6) 2025 DUE TO SHOULDER DISLOCATION. PREVIOUS SURGERY DATE IS UNKNOWN, AS WELL AS COMPLETE INFORMATION ON ORIGINAL PROSTHESIS. REVERSE LINER +3MM DIA40 (PART NUMBER 1365.50.815) HAS BEEN REMOVED, HOWEVER NO INDICATION OF WHICH MODEL OF LINER HAS BEEN IMPLANTED AS REPLACEMENT HAS BEEN PROVIDED. THE EVENT OCCURRED IN THE UNITED STATES.
Patient problems: Joint Dislocation
MAUDE report 24176349 ↗InjuryJan 23, 2026Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem
REVISION SURGERY WAS PERFORMED ON (B)(6) 2025 DUE TO SHOULDER DISLOCATION. PREVIOUS SURGERY IS UNKNOWN, AS WELL AS INFORMATION ABOUT ORIGINAL IMPLANT. DISLOCATION OCCURRED DUE TO SCAPULAR FRACTURE. SURGEON REMOVED THE REVERSE LINER (PART NUMBER 1365.50.815) AND REPLACED IT WITH A NEW RETENTIVE ONE (PART NUMBER 1365.50.816) AND THEN PROCEEDED WITH FIXING THE SCAPULA. SURGERY WAS COMPLETED AS INTENDED. PATIENT IS FEMALE, DATE OF BIRTH (B)(6) 1939. THE EVENT OCCURRED IN THE UNITED STATES.
Patient problems: Bone Fracture(s); Joint Dislocation
MAUDE report 24159381 ↗InjuryDec 12, 2025Unstable; Unstable
REVISION SURGERY PERFORMED ON (B)(6) 2025 DUE TO SHOULDER INSTABILITY. SURGEON REMOVED THE PRE-EXISTING SMR REVERSE LINER +3MM D.40MM (PART NUMBER: 1365.50.815, LOT: 23AT40L, STERILIZATION: 2400045) AND IMPLANTED AN EXTENSION FOR HUMERAL REVERSE BODY (PART NUMBER: 1352.15.001) AND A REVERSE LINER RETENTIVE STD DIA 40MM (PART NUMBER: 1365.50.811) TO IMPROVE SHOULDER STABILITY. PREVIOUS SURGERY DATE IS UNKNOWN, AS WELL AS OTHER COMPONENTS IMPLANTED IN PREVIOUS SURGERY. PATIENT IS A MALE, DATE OF BIRTH ON (B)(6) 1962. THE EVENT OCCURRED IN THE UNITED STATES.
Patient problems: Joint Laxity
MAUDE report 23789513 ↗InjuryAug 22, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
SHOULDER REVISION SURGERY WAS PERFORMED ON (B)(6) 2025, DUE TO DISLOCATION. THE SURGEON IMPLANTED A THICKER LINER. THEREFORE, THE ONLY COMPONENT REMOVED AND REPLACED BY A NEW ONE IS THE FOLLOWING: SMR REVERSE LINER +3MM D.40MM (PART CODE 1365.50.815, LOT NUMBER 22AT0NB, STERILIZATION (B)(4). THE OTHER COMPONENTS PREVIOUSLY IMPLANTED WERE LEFT IN SITU: TT AUGM.360 BASEPLATE #S-R+2MM (PART CODE 1375.15.522, LOT NUMBER 2211656, STERILIZATION (B)(4). SMR REVERSE HUMERAL BODY SHORT (PART CODE 1352.15.005, LOT NUMBER 2202509, STERILIZATION (B)(4). SMR GLENOSPHERE Ø 40MM (PART CODE 1374.09.121, LOT NUMBER 2206405, STERILIZATION (B)(4), SMR CONNECTOR SMALL R (PART CODE 1374.15.305, LOT NUMBER 2213086, STERILIZATION (B)(4), SMR GLENOID PEG TT SMALL-R #M (PART CODE 1375.14.652, LOT NUMBER 2213319, STERILIZATION (B)(4). SMR CEMENTLESS FINNED STEM (PART CODE 1304.15.180, LOT NUMBER 2209462, STERILIZATION (B)(4). PREVIOUS SURGERY TOOK PLACE ON (B)(6) 2022. THE PATIENT IS A FEMALE, DATE OF BIRTH (B)(6) 1951. THE EVENT HAPPENED IN THE UNITED STATES.
Patient problems: Joint Dislocation
MAUDE report 22872434 ↗InjuryJul 16, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
SHOULDER REVISION SURGERY WAS PERFORMED ON (B)(6) 2025, DUE TO DISLOCATION. THE FOLLOWING COMPONENTS PREVIOUSLY IMPLANTED WERE REMOVED AND REPLACED BY OTHER NEW COMPONENTS: SMR REVERSE LINER +3MM D.40MM (PART CODE: 1365.50.815, LOT NUMBER: 23AT40L, STERILIZATION: 2400045). CONNECTOR W. SCREW MED.LAT.7MM (PART CODE: 1974.15.302, LOT NUMBER: 2312609, STERILIZATION: 2300145). GLENOSPH.ECC. D.40MM ECC.4MM (PART CODE: 1976.09.040, LOT NUMBER: 2316007, STERILIZATION: 2300189). THE OTHER COMPONENTS PREVIOUSLY IMPLANTED WERE LEFT IN SITU: BASEPLATE D.25MM REGULAR (PART CODE: 1975.14.500, LOT NUMBER: 2405523, STERILIZATION: 2400083). SMR FINNED SHORT STEM D.21 (PART CODE: 1304.15.021, LOT NUMBER: 1917009, STERILIZATION: 2000019). SMR REVERSE HUMERAL BODY (PART CODE: 1352.15.010, LOT NUMBER: 2412894, STERILIZATION: 2400129). PREVIOUS SURGERY TOOK PLACE ON (B)(6) 2024. THE PATIENT IS A MALE, DATE OF BIRTH ON (B)(6) 1957. THE EVENT HAPPENED IN THE UNITED STATES.
Patient problems: Joint Dislocation
MAUDE report 22517532 ↗InjuryJun 19, 2025Device Dislodged or Dislocated; Patient Device Interaction Problem; Device Dislodged or Dislocated; Patient Device Interaction Problem
REVISION SURGERY PERFORMED ON (B)(6) 2025 DUE TO DISLOCATION. THE FOLLOWING COMPONENT PREVIOUSLY IMPLANTED, WAS REMOVED AND REPLACED BY A SMR REVERSE LINER RETENTIVE +6MM (PART CODE 1365.50.821): SMR REVERSE LINER +3MM D.40MM (PART CODE 1365.50.815, LOT NUMBER 24AT0US, STERILIZATION (B)(4). THE PATIENT UNDERWENT AN INITIAL TRAUMA SURGERY (PROXIMAL HUMERUS FRACTURE) ON (B)(6) 2025. THE OTHER DEVICES IMPLANTED ON THIS OCCASION WERE LEFT IN SITU DURING THE REVISION: SMR CEMENTLESS FINNED STEM (PART CODE 1304.15.170, LOT NUMBER 2427770, STERILIZATION (B)(4). SMR REVERSE FINNED HUMER. BODY (PART CODE 1352.15.050, LOT NUMBER 2310073, STERILIZATION (B)(4). SMR SMALL/STD CONNECTOR +2 (PART CODE 1374.15.322, LOT NUMBER 2427413, STERILIZATION (B)(4). SMR ECCENT. GLENOSPHERE DIA. 40MM (PART CODE 1376.09.041, LOT NUMBER 2433701, STERILIZATION (B)(4). AFTER THE REVISION SURGERY HEREBY REPORTED, THE PATIENT UNDERWENT ANOTHER REVISION PERFORMED ON (B)(6) 2025 (REPORTED BY COMPLAINT (B)(4), AND MFR. 3008021110-2025-00070). THE PATIENT IS A MALE, DATE OF BIRTH (B)(6) 1936. THE EVENT HAPPENED IN THE UNITED STATES.
Patient problems: Joint Dislocation
MAUDE report 22292117 ↗InjuryApr 4, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
REVISION SURGERY PERFORMED ON (B)(6) 2025, DUE TO IMPLANT DISLOCATION. PATIENT DISLOCATED TWICE. THE FOLLOWING LINER AND GLENOSPHERE WERE REMOVED AND THE SURGEON PUT IN A LATERALIZED GLENOSPHERE AND A LARGER RETENTIVE POLY: SMR REVERSE LINER +3MM D.40MM (PART CODE 1365.50.815, LOT NUMBER 23AT24A, STERILIZATION (B)(4), SMR ECCENT. GLENOSPHERE Ø 40MM (PART CODE 1376.09.041, LOT NUMBER 2307061, STERILIZATION (B)(4). PREVIOUS SURGERY TOOK PLACE ON (B)(6) 2025. THE PATIENT IS A MALE, DATE OF BIRTH (B)(6) 1955. THE EVENT HAPPENED IN THE UNITED STATES.
Patient problems: Joint Dislocation
MAUDE report 21770689 ↗
FDA clearance and classification
Cleared through 510(k) K142139 on Nov 7, 2014 (decision: Substantially Equivalent). Submitted by Lima Corporate S.P.A. as “Smr 40mm Glenosphere”. The FDA review took 94 days. FDA record ↗
The FDA regulates product code PHX (shoulder prosthesis, reverse configuration) under 21 CFR 888.3660 as a Class II device. 21 CFR 888.3660 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: Intended to be used to relieve pain and to replace the glenoid and humeral side of the shoulder joint in order to restore mobility in a grossly rotator cuff deficient joint with severe arthropathy or a previous joint replacement with a grossly rotator cuff deficient joint.
Common questions about the SMR Shoulder Liner
Has the SMR Shoulder Liner been recalled?
No FDA recall record in our data names this exact device identifier (DI 08033390102933), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the SMR Shoulder Liner?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between Apr 4, 2025 and Jul 2, 2026. In those reports, the most common specific device problems are “Device Dislodged or Dislocated” (7), “Patient Device Interaction Problem” (1) and “Unstable” (1); the most common patient problems are “Joint Dislocation” (7), “Bone Fracture(s)” (1) and “Joint Laxity” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SMR Shoulder Liner?
None of the reports are classified as deaths, 8 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the SMR Shoulder Liner FDA approved?
It was cleared through 510(k) K142139 on Nov 7, 2014. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the SMR Shoulder Liner MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the SMR Shoulder Liner, and is it still sold?
The FDA's device database lists Limacorporate SPA as the company responsible for it (brand SMR Shoulder, model Liner). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Shoulder Prosthesis, Reverse Configuration devices
- DJO SURGICAL 508-36-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration513 reports
- DJO SURGICAL 508-32-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration435 reports
- DJO SURGICAL 508-32-103Encore Medical, L.P. · shoulder prosthesis, reverse configuration427 reports
- DJO SURGICAL 509-00-036Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall258 reports
- DJO SURGICAL 508-32-104Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall202 reports
- DJO SURGICAL 509-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration239 reports
- DELTA XTEND 130760000Depuy (Ireland) · shoulder prosthesis, reverse configuration221 reports
- DJO SURGICAL 508-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration124 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page