Skip to content

SMR Shoulder Liner (REF 1365.50.811)Recalls & reported problems

Limacorporate SPA · Reverse Retentive Liner 40mm Standard · Class 2 · Implant

All models of SMR Shoulder (67)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033390102964), received between Jul 2, 2025 and Jul 2, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem1

Patient problems

  • Medical device site infection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 2, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    REVISION SURGERY WAS PERFORMED ON (B)(6) 2025. IT WAS REPORTED THAT THE PATIENT PRESENTED WITH A LARGE, UNEXPLAINED HEMATOMA. THE SURGEON SUSPECTED AN INFECTION; THEREFORE, THE WOUND WAS FLUSHED, AND THE LINER WAS SWAPPED. DURING THE PROCEDURE, THE SURGEON NOTED THAT THE IMPLANT WAS TOO LOOSE, SO THE VERSION WAS ADJUSTED, AND EXTENSIONS WERE ADDED. THE FOLLOWING SMR REVERSE FINNED HUMERAL BODY AND AN SMR REVERSE RETENTIVE LINER WERE REPLACED DURING THIS SURGERY: SMR REVERSE FINNED HUMER. BODY (PART CODE 1352.15.050, LOT NUMBER 2415343, STERILIZATION (B)(4)). SMR REVERSE LINER RETENTIVE (PART CODE 1365.50.811, LOT NUMBER 22AT145, STERILIZATION (B)(4) ). THE REVISION STEM SMR CEMENTL. REV. STEM Ø15 MM (PART CODE 1308.15.156, LOT NUMBER 1810713, STERILIZATION (B)(4)) REMAINED IN SITU. PREVIOUS SURGERY TOOK PLACE ON (B)(6) 2025. THE PATIENT IS A MALE, DATE OF BIRTH (B)(6) 1948. THE EVENT HAPPENED IN THE UNITED STATES.

    Patient problems: Medical device site infection

    MAUDE report 22381445 ↗

Other Shoulder Prosthesis, Reverse Configuration devices

All Shoulder Prosthesis, Reverse Configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page