BONE SCREWS 8431.15.040Recalls & reported problems

Limacorporate SPA · Cortical Bone Screw - Flattened Head Dia. 4,5 mm L 40 mm · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033390106870), received between Apr 27, 2018 and Apr 27, 2018 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1
  • Device Operates Differently Than Expected1
  • Loose or Intermittent Connection1

Patient problems

  • Inadequate Osseointegration1
  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 27, 2018Break; Loose or Intermittent Connection; Device Operates Differently Than Expected

    SMR REVERSE SHOULDER PROSTHESIS WAS IMPLANTED USING THE BONE GRAFT SURGICAL TECHNIQUE ON (B)(6) 2017. THE PROSTHESIS WAS REVISED ON (B)(6) 2018 DUE TO FAILURE OF THE BONE GRAFT (CAUSING LOOSENING OF THE METAL BACK + GLENOID PEG FROM GLENOID BONE) AND BREAKAGE OF ONE OF THE BONE SCREWS USED TO FIX THE METAL BACK GLENOID TO THE GLENOID BONE. THE PRODUCT CODE OF THE BONE SCREW IS NOT MARKETED IN THE US. THE PATIENT EXPERIENCED PAIN. THE SURGEON COMMENTED THAT THE BONE GRAFT USED MIGHT HAVE BEEN TOO LARGE. THE EVENT HAPPENED IN (B)(6).

    Patient problems: Inadequate Osseointegration; Pain

    MAUDE report 7467662 ↗

Other Shoulder Prosthesis, Reverse Configuration devices

All Shoulder Prosthesis, Reverse Configuration devices →