BONE SCREWS Self Tapping (REF 8420.15.005)Recalls & reported problems
Limacorporate SPA · Bone Screws dia. 6.5 h. 15mm · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033390110693), received between Nov 18, 2025 and Nov 18, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Break” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K172456); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Degraded1
- Patient Device Interaction Problem1
Patient problems
- Joint Laxity1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryNov 18, 2025Break; Degraded; Patient Device Interaction Problem; Break; Degraded; Patient Device Interaction Problem
SHOULDER REVISION SURGERY WAS PERFORMED ON (B)(6) 2025. DURING REVISION PROCEDURE, THE LINER WAS REMOVED AND EXAMINED (SMR REVERSE LINER STANDARD, PART CODE 1360.50.810, LOT NUMBER 20AT0XN, STERILIZATION (B)(4)), NOT SHOWING SIGNIFICANT OR UNUSUAL SIGNS OF WEAR. THE REVERSE BODY WAS LEFT IN PLACE BECAUSE IT WAS DETERMINED THAT REMOVING WOULD THREATEN TO FRACTURE TUBEROSITIES. GLENOSPHERE (SMR ECCENT. GLENOSPHERE Ø 36MM, PART CODE 1376.09.031, LOT NUMBER 2009277, STERILIZATION (B)(4)) AND BONE SCREWS (BONE SCREW Ø6,5 H.15MM, PART CODE 8420.15.005, LOT NUMBER 1802866, STERILIZATION (B)(4); BONE SCREW Ø6,5 H.20MM, PART CODE 8420.15.010, LOT NUMBER 2006407, STERILIZATION (B)(4); BONE SCREW Ø6,5 H.40MM, PART CODE 8420.15.050, LOT NUMBER 2006421, STERILIZATION (B)(4)) WERE REMOVED. AT THIS POINT, THE BASEPLATE (TT AUGM.360 BASEPLATE #S-R+2MM, PART CODE 1375.15.522, LOT NUMBER 2008603, STERILIZATION (B)(4)) COULD BE PULLED OUT BY HAND WITH HEMOSTATS. DURING REMOVAL OF SCREWS, IT WAS NOTED THAT THERE WERE METAL SHAVINGS (APPARENTLY FROM THE SCREWS) LEFT BEHIND THE BASEPLATE. IT WAS HYPOTHESIZED THAT THIS DEBRIS MAY HAVE BEEN CREATED WHILE IMPLANTING THE SCREWS AT TOO AGGRESSIVE OF AN ANGLE
Patient problems: Joint Laxity
MAUDE report 23579327 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K172456. One submission often covers many models, so check the product and lot details in each record.
- Z-0159-2022 · Oct 1, 2021 · TerminatedREF 8420.15.020, Bone Screw/Vite - Ti6AI4V, Self Tapping/Autofilettante, Dia=6.5mm, h=25mm, STERILE R, 2100099, UDI: (01)08033390018852
- Z-0160-2022 · Oct 1, 2021 · TerminatedREF 8420.15.010, Bone Screw/Vite - Ti6AI4V, Self Tapping/Autofilettante, Dia=6.5mm, h=20mm, STERILE R, 2100099, UDI: (01)08033390018845
Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.