Skip to content

SMR Shoulder Trial StemRecalls & reported problems

Limacorporate SPA · Trial Stem with Quick Connection Dia. 14 mm · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033390122115), received between Jan 2, 2026 and Jan 2, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K100858); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 2, 2026Break

    DURING REVERSE SHOULDER ARTHROPLASTY SURGERY PERFORMED ON (B)(6) 2025, WHILE BROACHING THE HUMERAL CANAL, THE QUICK CONNECT STEM IMPACTOR TIP (PART NUMBER 9013.02.142, LOT UNKNOWN) BROKE OFF INSIDE THE STEM TRIAL AS THE SURGEON WAS HITTING IT. TRIAL STEM WITH THE TIP OF THE BROACH HANDLE REMAINED STUCK INSIDE THE PATIENT AND COULD NOT BE REMOVED WITH STEM EXTRACTOR (PN 9013.02.301). MEDICAL TEAM USED OTHERS INSTRUMENTS AT THE FACILITY TO RETRIEVE THE BROKEN AND STUCK COMPONENTS. NO FRAGMENTS WERE LEFT IN PATIENT. SURGERY WAS PROLONGED OF ABOUT 20 MINUTES DUE TO REPORTED ISSUE BUT WAS COMPLETED WITHOUT FURTHER COMPLICATIONS. NO IMPACT ON PATIENT HAS BEEN REPORTED. PATIENT IS FEMALE, DATE OF BIRTH (B)(6) 1962. THE EVENT OCCURRED IN THE UNITED STATES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23944366 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K100858. One submission often covers many models, so check the product and lot details in each record.

  • Z-2539-2015 · Aug 14, 2015 · Terminated
    SMR Shoulder/Spalla Trial Humeral Head diameter .42 mm, REF 9013.22.425; SMR Shoulder/Spalla Trial Humeral Head diameter .44 mm, REF 9013.22.445; SMR Shoulder/Spalla Trial Humeral
  • Z-2540-2015 · Aug 14, 2015 · Terminated
    SMR Shoulder/Spalla Trial Heads ECC .4 mm Adaptor, REF 9013.30.021; SMR Shoulder/Spalla Trial Heads ECC .8 mm Adaptor, REF 9013.30.031; for use in shoulder implant surgery.

Other Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented devices

All Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.