Skip to content

SMR Shoulder 9013.52.147Recalls & reported problems

Limacorporate SPA · Reverse Adaptor Sleeve · Class 2

All models of SMR Shoulder (67)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08033390152068), received between Jan 7, 2025 and May 5, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

2 other recalls cover devices cleared under the same FDA submission (K100858); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 5, 2025Break; Break

    AFTER A SHOULDER SURGERY PERFORMED ON (B)(6) 2025, THE OPERATIONS COORDINATOR IN THE SUBSIDIARY'S WAREHOUSE WAS CHECKING THE INSTRUMENT SET THAT WAS RETURNED FROM THE HOSPITAL AND FOUND THAT THE TIP OF THE REVERSE ADAPTOR SLEEVE (PRODUCT CODE 9013.52.147, LOT #22BH0AK) WAS BROKEN. THERE IS NO INFORMATION AS TO HOW THE DEVICE BROKE. STILL, NO CONSEQUENCES WERE REPORTED. EVENT HAPPENED IN AUSTRALIA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21960276 ↗
  • MalfunctionJan 7, 2025Break; Break

    DURING A PRIMARY SURGERY PERFORMED ON (B)(6) 2024, THE FRONT PART OF THE REVERSE ADAPTOR SLEEVE (PRODUCT CODE: 9013.52.147, LOT#: 23BH002) FOR THE CONNECTION WITH THE HUMERAL BODY TRIAL IS BROKEN WHEN THE SURGEON TRIED TO TAKE OUT THE TRIAL FROM THE HUMERAL CAVITY. DUE TO THE BROKEN INSTRUMENT, IT WAS DIFFICULT TO TAKE OUT THE TRIAL FROM THE HUMERAL CAVITY. NO OTHER INSTRUMENT WAS AVAILABLE: THE SURGEON HELD THE TRIAL WITH A RONGEUR AND KNOCKED IT OUT CAREFULLY. IT WAS REPORTED THAT THE INSTRUMENT WAS USED ABOUT 10 TIMES. THE SURGERY GOT PROLONGED FOR 40 MINUTES. PATIENT IS A FEMALE, 61 YEARS OLD. EVENT HAPPENED IN CHINA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21091068 ↗
  • MalfunctionJan 7, 2025Break; Break

    DURING PRIMARY SHOULDER SURGERY PERFORMED ON (B)(6) 2024 THE FRONT PART REVERSE ADAPTOR SLEEVE (PRODUCT CODE 9013.52.147, LOT #2117361) FOR THE CONNECTION WITH THE HUMERAL BODY TRIAL GETS BROKEN WHEN THE SURGEON TRIED TO TAKE OUT THE TRIAL FROM THE HUMERAL CAVITY. DUE TO THE BROKEN INSTRUMENT, IT WAS DIFFICULT TO TAKE OUT THE TRIAL FROM THE HUMERAL CAVITY. NO OTHER INSTRUMENT WAS AVAILABLE: THE SURGEON HELD THE TRIAL WITH A RONGEUR AND KNOCKED IT OUT CAREFULLY. ACCORDING TO THE RECEIVED INFORMATION, THE INSTRUMENT WAS USED ABOUT 20 TIMES. THE SURGERY GOT PROLONGED FOR 40 MINUTES. PATIENT IS A MALE, 71 YEARS OLD. EVENT HAPPENED IN CHINA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21091067 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K100858. One submission often covers many models, so check the product and lot details in each record.

  • Z-2540-2015 · Aug 14, 2015 · Terminated
    SMR Shoulder/Spalla Trial Heads ECC .4 mm Adaptor, REF 9013.30.021; SMR Shoulder/Spalla Trial Heads ECC .8 mm Adaptor, REF 9013.30.031; for use in shoulder implant surgery.
  • Z-2539-2015 · Aug 14, 2015 · Terminated
    SMR Shoulder/Spalla Trial Humeral Head diameter .42 mm, REF 9013.22.425; SMR Shoulder/Spalla Trial Humeral Head diameter .44 mm, REF 9013.22.445; SMR Shoulder/Spalla Trial Humeral

Other Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented devices

All Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page