Skip to content

Minima S Instrument Set 9045.03.207Recalls & reported problems

Limacorporate SPA · Trial Modular Neck #7 · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033390153928), received between Aug 22, 2024 and Aug 22, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Mechanical Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 22, 2024Mechanical Problem; Mechanical Problem

    DURING THE TEST REDUCTION OF A PRIMARY HIP ARTHROPLASTY PERFORMED ON (B)(6) 2024 , AN UNUSUAL NOISE WAS MADE AS IF THE LINER TRIAL AND HEAD TRIAL WERE RUBBING AGAINST EACH OTHER. WHEN DISLOCATING, IT WAS FOUND THAT THE MINIMA S TRIAL MODULAR NECK #7 (PRODUCT CODE 9045.03.207, LOT #17HM0AC) WAS INSERTED INTO THE BROACH TO THE POINT THAT THERE WAS NO GAP BETWEEN THEM. THIS SHORTENED THE TRIAL NECK LENGTH BY ABOUT 2 MM, MAKING IT DIFFICULT TO CONFIRM THE EXACT SOFT TISSUE BALANCE. BECAUSE THERE WAS NO GAP, THE NECK EXTRACTOR COULDN'T BE INSERTED, AND IT WAS MANAGED TO LIFT THE NECK TRIAL USING THE HOSPITAL'S K-WIRE AND INSERTING A ZUK REMOVAL HOOK TO REMOVE IT. THE SURGERY WAS PROLONGED FOR ABOUT 20 MINUTES DUE TO THE DIFFICULTIES ENCOUNTERED IN REMOVING THE NECK TRIAL. PATIENT IS A MALE, 74 YEARS OLD. IT WAS REPORTED HE'S A LARGE MALE WITH AMPLE MUSCLE MASS. EVENT HAPPENED IN JAPAN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20036627 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page