Skip to content

BIOLOX DELTA 5010.42.323Recalls & reported problems

Limacorporate SPA · Biolox Delta - Femoral Head Dia 32 mm #L · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033390179904), received between Nov 20, 2025 and Nov 20, 2025 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Medical device site infection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HIP REVISION SURGERY DUE TO INFECTION PERFORMED ON (B)(6) 2025. COMPLAINT SOURCE REPORTED APPEARANCE OF HYPEREMIC TUMESCENCE WITH FISTULATION ON THE SCAR UPON ROUTINE CHECKUP, COMPATIBLE WITH ACUTE SEPTIC PHLEBITIS. EXPLANTATION, MULTIPLE SWABS AND APPLICATION OF ANTIBIOTIC SPACER. PREVIOUS SURGERY WAS PERFORMED ON (B)(6) 2025 (DUE TO LEFT FEMORAL NECK FRACTURE). THE FOLLOWING DEVICES WERE EXPLANTED: DELTA PROTR.LINER ØINT 32MM #M, PRODUCT CODE 5886.54.158, LOT N. 25AT118, STER. N. 2500100 FEM. MODULAR HEAD - L Ø32MM, PRODUCT CODE 5010.42.323, LOT N. 7011917337 PATIENT IS FEMALE, 48 YEARS OLD. THIS EVENT OCCURRED IN ITALY. NOTE: ONLY THE HEAD IS CLEARED IN THE USA.

    Patient problems: Medical device site infection

    MAUDE report 23605997 ↗

Other Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous devices

All Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page