Skip to content

AMF REVISION TT CONES 906C.21.818Recalls & reported problems

Limacorporate SPA · AMF Revision TT Cones - Central Tibia Reamer Head Dia. 18 mm · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033390189569), received between Jul 15, 2024 and Jul 15, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 15, 2024Break; Break

    INTRA-OPERATIVE ISSUE DUE TO BREAKAGE HAPPENED ON (B)(6) 2024. IT WAS NOT THE FIRST TIME IT WAS USED; THE CONSIGNMENT SET INVOLVED HAS BEEN USED PROBABLY 5-10 TIMES. THE INSTRUMENTS INVOLVED IN THE EVENT ARE THE FOLLOWING: REAMING SHAFT (PART CODE 906C.10.800, LOT NUMBER 22BQ01Y). CENTRAL TIBIA REAMER HEAD D.18 (PART CODE 906C.21.818, LOT 22BG037). THERE WAS NO IMPACT ON SURGERY (ABOUT 10 MINUTES DELAY) AND THE SURGEON WAS ABLE TO WORK AROUND IT, BECAUSE WHEN THE SHAFT WENT THROUGH THE REAMER, HE HAD COMPLETED REAMING AND WAS ABLE TO PLACE THE IMPLANT. THE PATIENT IS A FEMALE, DATE OF BIRTH (B)(6) 1952. THE EVENT HAPPENED IN THE UNITED STATES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19746262 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page