Skip to content

AMF TT CONES 6C21.14.180Recalls & reported problems

Limacorporate SPA · AMF TT CONES - Central Tibia Dia. 18 mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033390230919), received between Jul 22, 2026 and Jul 22, 2026 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Medical device site infection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 22, 2026Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY PERFORMED ON (B)(6) 2026 DUE TO INFECTION. PREVIOUS SURGERY WAS ON (B)(6) 2026 WHEN ANOTHER MANUFACTURER KNEE IMPLANTS USED WITH THE AMF TT CONES: CONE CENTRAL FEMUR D.18 (PN 6C10.14.180, LOT 2401822 STERILIZATION (B)(4)). CONE CENTRAL TIBIA D.18 (PS 6C21.14.180, 2536933 STERILIZATION (B)(4)). AFTER FEW MONTHS PATIENT DEVELOPED AN INFECTION AND THE PRE-EXISTING AMF TT CONES WERE REPLACED WITH NEW ONES WITH DIA 21MM (PN 6C10.14.210 AND 6C21.14.210 RESPECTIVELY). THE WHOLE IMPLANT FROM THE THIRD-PARTY MANUFACTURER WAS ALSO REPLACED. SURGERY WAS COMPLETED AS INTENDED. PATIENT IS MALE, DATE OF BIRTH (B)(6) 1960. THE EVENT OCCURRED IN THE UNITED STATES.

    Patient problems: Medical device site infection

    MAUDE report 25903356 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page