Skip to content

PRIMA TT Glenoid 9019.74.301Recalls & reported problems

Limacorporate SPA · PRIMA TT Glenoid - Connector Positioner/Extractor · Class 2

All models of PRIMA TT Glenoid (16)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033390257909), received between Jan 7, 2025 and Jan 7, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Mechanical Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 7, 2025Mechanical Problem; Mechanical Problem

    INTRA-OPERATIVE ISSUE OCCURRED DURING A PRIMA TT IMPLANT SURGERY PERFORMED ON (B)(6) 2024. AFTER IMPLANTING A GLENOSPHERE TRIAL 36 MEDIUM, THE SURGEON WENT THROUGH RANGE OF MOTION AND WAS NOT HAPPY. HE DECIDED TO REMOVE GLENOSPHERE TO TRIAL A 40 MEDIUM. AFTER REMOVING THE SET SCREW FOR GLENOSPHERE, HE INSERTED THE TAPER BREAKER OF THE FOLLOWING INSTRUMENT: CONNECTOR POSITIONER/EXTRACTOR (PART CODE 9019.74.301, LOT NUMBER 24BH00Q). THE NARROW ROD FOR TAPER BREAKER GOT STUCK ON THE IMPLANT AND COULD NOT BE REMOVED. HE HAD TO DISASSEMBLE THE TAPER BREAKER TO REMOVE THE GLENOSPHERE AND ROD IN SEPARATE PIECES. AFTER REMOVING THE GLENOSPHERE, IT WAS NOTED THAT THE MEDIUM CONNECTOR WAS LEFT BEHIND. THE SURGEON ATTEMPTED TO USE THE NEW REMOVAL DEVICE FROM THE TRAY, BUT EVERY TIME IT WAS LOCKED IN PLACE THE BREAKER ROD WAS INSERTED, THE DEVICE DISCONNECTED FROM THE IMPLANT. THIS WAS ATTEMPTED SEVERAL TIMES WITH THE SAME RESULT. EVENTUALLY, AFTER WASTING ABOUT 10 MINUTES OF OR TIME, THE SURGEON DECIDED TO IMPLANT A 40 GLENOSPHERE ON TOP OF THE MEDIUM CONNECTOR. TAPER DID NOT APPEAR TO BE DAMAGED ON THE INSIDE, SO HE FELT COMFORTABLE IMPLANTING A NEW GLENOSPHERE INTO IT. THE PATIENT IS A MALE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21090855 ↗

Other Shoulder Prosthesis, Reverse Configuration devices

All Shoulder Prosthesis, Reverse Configuration devices →

Sources and updates