SMR System 1374.09.105Recalls & reported problems
Limacorporate SPA · SMR System - Glenosphere Dia. 36 mm Small-R · Class 2 · Implant
All models of SMR System (47)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08033390272216), received between May 17, 2023 and Nov 11, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
1 other recall cover devices cleared under the same FDA submission (K110598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Device Dislodged or Dislocated1
- Unstable1
Patient problems
- Insufficient Information1
- Joint Dislocation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
DURING A SURGERY ON (B)(6) 2025, THE SURGEON IMPLANTED THE METAL-BACK GLENOID SMALL-R (PRODUCT CODE: 1375.20.005 - LOT: 2513700 - STER: (B)(4) AND FIXED IT WITH TWO SCREWS, AS PLANNED. AFTER THAT, THE 36 GLENOSPHERE (PRODUCT CODE: 1374.09.105 - LOT: 2514000 - STER: (B)(4) WAS CONNECTED TO THE CONNECTOR FROM THE SAME PACKAGE AND ATTACHED TO THE IMPACTOR, WHICH WAS FULLY SECURED. THE ASSEMBLY WAS PLACED IN SITU USING A MALLET. HOWEVER, THE GLENOSPHERE COULD NOT BE FIXED. CONSEQUENTLY, THE SURGEON REMOVED IT, DISCONNECTED THE CONNECTOR FROM THE GLENOSPHERE, AND FIXED IT TO THE METAL-BACK. A FURTHER ATTEMPT TO PLACE THE GLENOSPHERE WAS MADE, BUT AGAIN IT WAS NOT POSSIBLE. THE SCREWS, THE METAL-BACK, AND THE CONNECTOR WERE REMOVED TOGETHER. THE CONNECTOR WAS THEN DETACHED FROM THE METAL-BACK, AND THE IMPLANTATION PROCESS WAS RESTARTED, YET THE SAME ISSUE OCCURRED. AS A RESULT, THE SURGEON REMOVED THE COMPONENTS AND CHANGED TO A METAL-BACK SMALL AND THE SUITABLE GLENOSPHERE 36 AND CONNECTOR. THERE WAS SUFFICIENT BONE STOCK AT THE GLENOID, ALLOWING THE IMPLANTS TO BE PLACED IN SITU WITHOUT ANY FURTHER ISSUES. THE PHYSICIAN WHO PERFORMED THE SURGERY IS AN SMR-EXPERIENCED SURGEON. THERE WAS
Patient problems: Insufficient Information
MAUDE report 23523309 ↗InjuryMay 17, 2023Unstable; Device Dislodged or Dislocated; Unstable; Device Dislodged or Dislocated
SHOULDER REVISION SURGERY PERFORMED ON (B)(6) 2018: ROUTINE POST OP X-RAYS REVEALED A SUBLUXED HUMERAL PROSTHESIS. THE GLENOSPHERE WAS UPSIZED TO A 40 MM GLENOSPHERE. DURING THIS REVISION SURGERY, THE FOLLOWING DEVICES WERE EXPLANTED: - SMR REV. LINER RETENTIVE +6 MM CODE 136150020, LOT 15AT1H5. - SMR GLENOSPHERE Ø36MM SMALL-R CODE 137409105, LOT 1713817. THE PREVIOUS SURGERY WAS CARRIED OUT ON (B)(6) 2018 AND IT WAS A REVISION SURGERY. (REGISTERED AS COMPLAINT 153/23, NOT REPORTED TO FDA AS THE INVOLVED PRODUCT IS NOT MARKETED IN THE US). THE COMPLAINT SOURCE REPORTED POOR MUSCLE CONTROL AFTER CEMENT SPACER FOR SEVERAL MONTHS AND INSTABILITY WITH THE 36 MM LONG RETENTIVE LINER. THIS EVENT OCCURRED IN THE UNITED KINGDOM. NOTE: THE PRODUCT CODE 136150020 IS NOT MARKETED IN THE US.
Patient problems: Joint Dislocation
MAUDE report 16945189 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K110598. One submission often covers many models, so check the product and lot details in each record.
- Z-1711-2016 · Apr 8, 2016 · TerminatedSMR glenosphere impactor/extractor; Used to implant the glenosphere in the SMR Reverse Shoulder replacement.
Other Shoulder Prosthesis, Reverse Configuration devices
- DJO SURGICAL 508-36-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration508 reports
- DJO SURGICAL 508-32-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration433 reports
- DJO SURGICAL 508-32-103Encore Medical, L.P. · shoulder prosthesis, reverse configuration421 reports
- DJO SURGICAL 509-00-036Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall259 reports
- DJO SURGICAL 508-32-104Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall203 reports
- DJO SURGICAL 509-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration233 reports
- DELTA XTEND 130760000Depuy (Ireland) · shoulder prosthesis, reverse configuration223 reports
- DJO SURGICAL 508-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration122 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page