Skip to content

SMR System 1350.15.030Recalls & reported problems

Limacorporate SPA · SMR System - Humeral Body Trauma Short · Class 2 · Implant

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033390272506), received between Mar 17, 2025 and Mar 17, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K100858); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem1

Patient problems

  • Inadequate Osseointegration1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 17, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    REVISION SURGERY PERFORMED ON (B)(6) 2025, DUE TO GLENOID LOOSENING, BONE LOSS AND POOR BONE QUALITY. THE FOLLOWING COMPONENTS WERE REMOVED: SMR TRAUMA HUM. BODY# SHORT (PART NUMBER: 1350.15.030, LOT NUMBER: 2017188, STERILIZATION: 2000339). SMR HUMERAL HEAD Ø42 MM (PART NUMBER: 1322.09.420, LOT NUMBER: 1400754, STERILIZATION: 2100055). SMR ECC. ADAPTOR TAPER STANDARD (PART NUMBER: 1330.15.274, LOT NUMBER: 2124614, STERILIZATION: 2200002). THE PATIENT PREVIOUSLY HAD A TOTAL SHOULDER ARTHROPLASTY WITH LIMA COMPONENTS (SMR STEM, TRAUMA BODY, ECCENTRIC ADAPTER, HUMERAL HEAD, AND 3-PEGGED CEMENTED POLY), THOUGH THE INITIAL SURGERY DATE IS UNKNOWN. ON (B)(6) 2024, THE PATIENT UNDERWENT A REVISION PROCEDURE FOR GLENOID LOOSENING. DURING THE ON (B)(6) SURGERY, THE SURGEON ATTEMPTED TO COMPLETE THE SECOND PHASE OF THE REVISION. THE HUMERAL STEM, BODY, AND HEAD WERE REMOVED. HOWEVER, WHEN TRYING TO IMPLANT THE BASEPLATE FOR A REVERSE SHOULDER REPLACEMENT, THE SOFT GLENOID BONE CRACKED. TO ADDRESS THIS, THE SURGEON DECIDED TO REINFORCE THE GLENOID WITH BONE GRAFT AND PROCEEDED BY PLACING AN ADAPTER AND HUMERAL HEAD ONTO THE REVERSE BODY, CONVERTING IT INTO A HEMIARTHROPLASTY. THE PATIENT IS

    Patient problems: Inadequate Osseointegration

    MAUDE report 21620230 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K100858. One submission often covers many models, so check the product and lot details in each record.

  • Z-2539-2015 · Aug 14, 2015 · Terminated
    SMR Shoulder/Spalla Trial Humeral Head diameter .42 mm, REF 9013.22.425; SMR Shoulder/Spalla Trial Humeral Head diameter .44 mm, REF 9013.22.445; SMR Shoulder/Spalla Trial Humeral
  • Z-2540-2015 · Aug 14, 2015 · Terminated
    SMR Shoulder/Spalla Trial Heads ECC .4 mm Adaptor, REF 9013.30.021; SMR Shoulder/Spalla Trial Heads ECC .8 mm Adaptor, REF 9013.30.031; for use in shoulder implant surgery.

Other Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented devices

All Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.