Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033477071909), received between May 29, 2019 and May 29, 2019 — classified as 1 injury.

The most frequently reported specific device problem is “Deflation Problem” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K092177); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Deflation Problem1

Patient problems

  • Patient Problem/Medical Problem1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 29, 2019Deflation Problem; Deflation Problem

    IT WAS REPORTED THAT THE PHYSICIAN WAS USING A MOMA ULTRA DEVICE IN THE TREATMENT OF A PLAQUE LESION IN THE LEFT PROXIMAL COMMON CAROTID ARTERY(CCA). THE DEVICE WAS PREPPED AS PER IFU WITH NO ISSUES IDENTIFIED. THE PROXIMAL CCA BALLOON WOULD NOT COMPLETELY DEFLATE EVEN WITH 50/50 SALINE CONTRAST PEP. A 60CC SYRINGE OF NEGATIVE PRESSURE WAS NEEDED TO DEFLATE THE BALLOON. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.

    Patient problems: Patient Problem/Medical Problem

    MAUDE report 8649607 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K092177. One submission often covers many models, so check the product and lot details in each record.

  • Z-0320-2024 · Oct 12, 2023 · Open, Classified
    Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MUS0130069X6
  • Z-0321-2024 · Oct 12, 2023 · Open, Classified
    Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MOM0130069X6
  • Z-0322-2024 · Oct 12, 2023 · Open, Classified
    Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MOM0130068X5

Other Temporary Carotid Catheter for Embolic Capture devices

All Temporary Carotid Catheter for Embolic Capture devices →