Skip to content

PACIFIC XTREME PCU050200130 (UDI 08033477076065)Recalls & reported problems

Medtronic, Inc. · CATH PCU050200130 PCU US 05 00L200UL1300 · Class 2

All models of PACIFIC XTREME (25)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08033477076065), received between Jun 4, 2020 and Mar 27, 2021 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K103464); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Leak/Splash2
  • Material Twisted/Bent2
  • Burst Container or Vessel1
  • Inflation Problem1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 27, 2021Burst Container or Vessel; Leak/Splash; Material Twisted/Bent; Burst Container or Vessel; Leak/Splash; Material Twisted/Bent

    PHYSICIAN WAS ATTEMPTING TO USE PACIFIC XTREME PTA BALLOON DURING PROCEDURE TO TREAT A MODERATELY CALCIFIED LESION IN THE RIGHT MID TO DISTAL SUPERFICIAL FEMORAL ARTERY (SFA) WITH 50% STENOSIS. THE VESSEL WAS NONE TORTUOUS. THE VESSEL DIAMETER AND LESION LENGTH ARE 5MM AND 170MM RESPECTIVELY. A MEDTRONIC INFLATION DEVICE WAS USED. THERE WAS NO DAMAGE NOTED TO PACKAGING. THERE WAS NO ISSUE NOTED WHEN REMOVING THE DEVICE FROM HOOP/TRAY. THE DEVICE WAS NOT PREPPED PER IFU WITH NO ISSUES IDENTIFIED. DURING BALLOON INFLATION A PINHOLE LEAK WAS NOTED AT 8ATM. ALL BALLOON FRAGMENTS WERE RETRIEVED. BALLOON TWIST WAS ALSO NOTED. THE DEVICE DID NOT PASS THROUGHA PREVIOUSLY DEPLOYED STENT. THERE WAS NO RESISTANCE ENCOUNTERED WHEN ADVANCING THE DEVICE. IT WAS REPORTED THAT THE BALLOON WAS ADVANCED AND UPON INFLATION, UNDEPLOYED SEGMENT WAS NOTED UNDER FLUOROSCOPY. BALLOON WAS DEFLATED AND REINFLATED BUT UNDEPLOYED SEGMENT STILL VISIBLE. DURING SECOND INFLATION BALLOON BURST AT NOMINAL PRESSURE. BALLOON WAS REMOVED AND ANOTHER BALLOON WAS USED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11579997 ↗
  • MalfunctionOct 9, 2020Inflation Problem; Material Twisted/Bent; Inflation Problem; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN ATTEMPT TO USE PACIFIC XTREME BALLOONS, THE TWO BALLOONS WERE BOTH INSERTED INTO THE BODY, AND WHEN INFLATED DID NOT FULLY INFLATE IN THE MIDDLE OF THE BALLOON.

    Patient problems: Insufficient Information; No Information

    MAUDE report 10655042 ↗
  • MalfunctionJun 4, 2020Leak/Splash; Leak/Splash

    PHYSICIAN WAS ATTEMPTING TO USE A PACIFIC XTREME PTA BALLOON CATHETER ALONG WITH NON-MEDTRONIC 6FR SHEATH AND 0.014 GUIDE WIRE DURING PROCEDURE TO TREAT A SEVERELY CALCIFIED LESION IN THE RIGHT MID SFA WITH CHRONIC TOTAL OCCLUSION (CTO-100%). THE VESSEL WAS SEVERELY TORTUOUS WITH 6MM DIAMETER. THERE WERE NO DAMAGES TO THE PACKAGINGS. THERE WERE NO ISSUES NOTED WHEN REMOVING DEVICES FROM HOOP/TRAY. THE DEVICES WERE PREPPED PER IFU WITH NO ISSUES IDENTIFIED. THE BALLOONS WERE INFLATED USING AN INFLATION DEVICE. THE DEVICES PASSED THROUGH A PREVIOUSLY DEPLOYED STENT WITH NO RESISTANCE FELT WHEN ADVANCING THE DEVICE. DURING PACIFIC XTREME BALLOON INFLATION, PIN HOLE LEAK WAS NOTED AT 8ATM. PHYSICIAN EXCHANGED FOR A FORTREX PTA BALLOON AND DURING BALLOON INFLATION, A PIN HOLE LEAK WAS ALSO NOTED AT 8ATM. IT WAS REPORTED THAT THE FIRST BALLOON RUPTURED AT THE DISTAL END PRIOR STENT IT WAS A 5X200X130 PACIFIC XTREME .018 BALLOON. THEN A 6X40X135 FORTREX BALLOON WAS USED AND ALSO RUPTURED. THEN A 6X60X120 EVERFLEX STENT WAS PLACED AT DISTAL END OF PREVIOUSLY DEPLOYED STENT. THEN A 6X20X135 FORTREX BALLOON WAS USED TO POST DILATE AND THE PROCEDURE WAS COMPLETED. THERE WAS NO PATIENT INJURY

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10117655 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K103464. One submission often covers many models, so check the product and lot details in each record.

  • Z-3055-2020 · Aug 26, 2020 · Terminated
    Pacific Xtreme (PTCA Balloon Dilation Catheter)

Other Catheter, percutaneous devices

All Catheter, percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page