Empower H65 PVMS00051Recalls & reported problems

Quanta System SPA · Distributed by Gyrus ACMI, L.P. · Class 2

All models of Empower H65 (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Quanta Empower H65 is a powered laser surgical instrument distributed by Gyrus ACMI. The 59 reports mostly describe the laser failing to deliver output or shutting down mid-procedure, sometimes from overheating, which delayed treatment; no reports describe injuries or deaths. No recalls name this device, and reports have not been filed in the past 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
59
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 59 adverse event reports that list this exact device (DI 08033945938192), received between Nov 6, 2018 and Nov 28, 2022 — classified as 59 malfunctions.

The most frequently reported specific device problem is “Therapeutic or Diagnostic Output Failure” (26 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Therapeutic or Diagnostic Output Failure26
  • Output Problem14
  • Defective Device7
  • Appropriate Device Problem Term/Code Not Available5
  • Electrical /Electronic Property Problem3
  • Complete Loss of Power1
  • Defective Component1
  • Fluid/Blood Leak1

Patient problems

  • No Known Impact Or Consequence To Patient31
  • No Clinical Signs, Symptoms or Conditions28

Recent adverse event reports

The 25 most recent of 59 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 28, 2022Output Problem; Output Problem

    THE DEVICE HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15868562 ↗
  • MalfunctionNov 28, 2022Output Problem; Output Problem

    THE DEVICE HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15868551 ↗
  • MalfunctionNov 4, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW THE PATIENT TO USE IT. NO ADVERSE EFFECTS TO PATIENT HAVE BEEN REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15728055 ↗
  • MalfunctionSep 14, 2022Output Problem; Output Problem

    LASER OVERHEATING AND SHUTTING DOWN DURING SURGICAL PROCEDURE CAUSING TREATMENT DELAY. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15415830 ↗
  • MalfunctionJun 30, 2022Defective Component

    THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14869716 ↗
  • MalfunctionJun 23, 2022Defective Device

    THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14785549 ↗
  • MalfunctionJun 16, 2022Optical Problem; Optical Problem

    THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14711267 ↗
  • MalfunctionFeb 24, 2022Therapeutic or Diagnostic Output Failure

    THE LASER HAD A FAILURE THAT DID NOT ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13590668 ↗
  • MalfunctionFeb 4, 2022Therapeutic or Diagnostic Output Failure

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13451390 ↗
  • MalfunctionJan 25, 2022Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13342784 ↗
  • MalfunctionJan 25, 2022Infusion or Flow Problem; Output Problem; Infusion or Flow Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13342773 ↗
  • MalfunctionDec 16, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13017691 ↗
  • MalfunctionNov 16, 2021Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12818367 ↗
  • MalfunctionOct 22, 2021No Display/Image; No Display/Image

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY.NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12679541 ↗
  • MalfunctionOct 7, 2021Radiation Output Problem; Radiation Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12588642 ↗
  • MalfunctionAug 27, 2021Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12380364 ↗
  • MalfunctionAug 13, 2021Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12317555 ↗
  • MalfunctionAug 12, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12311454 ↗
  • MalfunctionJul 23, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12221194 ↗
  • MalfunctionJul 4, 2021Output Problem; Complete Loss of Power; Output Problem; Complete Loss of Power

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12114600 ↗
  • MalfunctionJun 3, 2021Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11930572 ↗
  • MalfunctionMay 11, 2021Therapeutic or Diagnostic Output Failure

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11805680 ↗
  • MalfunctionFeb 12, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11320364 ↗
  • MalfunctionJan 8, 2021Therapeutic or Diagnostic Output Failure

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11138008 ↗
  • MalfunctionOct 22, 2020Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10719036 ↗