Single use Surgical Optical Fiber OFH302721Recalls & reported problems
Quanta System SPA · Single use Ball tip Surgical Optical Fiber ø 272µm. Distributed by Olympus Surgical Technologies America.Gyrus ACMI, Inc. · Class 2
All models of Single use Surgical Optical Fiber (18)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08033945938284), received between Jan 28, 2020 and May 22, 2020 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K180158); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break2
- Therapeutic or Diagnostic Output Failure1
- Thermal Decomposition of Device1
Patient problems
- No Known Impact Or Consequence To Patient2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 22, 2020Therapeutic or Diagnostic Output Failure
THE OPTICAL FIBER HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10082907 ↗MalfunctionFeb 3, 2020Break
DURING LASER LITHOTRIPSY THE EMPOWER LASER FIBER WIRE SNAPPED. IT BROKE AT THE ENTRY POINT OF THE FLEXIBLE URETEROSCOPE. ANOTHER FIBER WIRE WAS SUPPLIED AND THE EXACT EVENT HAPPENED AT THE SAME POINT OF ENTRY. IT WAS NOTICED THAT BOTH FIBER WIRES WERE THE SAME REFERENCE AND LOT NUMBER. A THIRD LASER FIBER WIRE WAS SUPPLIED THAT WAS THE SAME 200 SERIES BUT A DIFFERENT REFERENCE AND LOT NUMBER. IT DID NOT SNAP THOUGH IT WAS NOTICED THAT THE FIBER WIRE WAS NOT LIGHTING UP WITH THE GREEN BEAM AT THE END OF THE FIBER AS IT SHOULD. CHARGE RN NOTIFIED THAT ALL 200 SERIES WIRES WILL BE GONE THROUGH AND ANY WITH THE SAME REFERENCE AND LOT NUMBER AS THE DEFECTIVE FIBERS AND WILL BE PLACED IN MATERIAL BIN WITH A NOTE AS TO WHY. THE EFFECTED WIRES FROM THIS EVENT WERE SENT TO CLINICAL ENGINEERING AND EXAMINED UNDER A MICROSCOPE. THE LASER FIBERS WILL BE RETURNED TO THE MANUFACTURER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9659279 ↗MalfunctionJan 28, 2020Break; Thermal Decomposition of Device
EMPOWER STERILE SURGICAL LASER FIBER BROKE IN HALF WHERE IT ENTERED THE FLEXIBLE URETEROSCOPY SCOPE. AFTER IT BROKE IT LANDED ON THE DRAPE AND BURNED A TINY HOLE IN DRAPE. NO HARM WAS CAUSED TO THE PATIENT. NO HARM WAS CAUSED TO ANY OF THE STAFF. BOTH PIECES OF THE LASER FIBER WERE KEPT AND TURNED IN. FIBER WILL BE RETURNED BY SITE STAFF FOR FAILURE ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9636380 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K180158. One submission often covers many models, so check the product and lot details in each record.
- Z-1842-2021 · Apr 27, 2021 · TerminatedOlympus Sterile Optical Laser Fiber, single use and reusable
Other Powered Laser Surgical Instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page