VenaCure 1470 Pro PDMS00252Recalls & reported problems

Quanta System SPA · Distributed by: Angiodynamics Inc. 603 Queensbury Ave., NY 12804 USA · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This VenaCure 1470 Pro surgical laser is used for vein procedures. Most of the 42 reports that name its identifier describe the laser failing to produce output or power up, so treatment could not proceed, with no patient effects noted; 1 injury report mentions blood clots found at follow-up after treatment. There were no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
42
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 42 adverse event reports that list this exact device (DI 08033945938925), received between Oct 30, 2019 and Jun 11, 2024 — classified as 1 injury, 41 malfunctions.

The most frequently reported specific device problem is “Therapeutic or Diagnostic Output Failure” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Therapeutic or Diagnostic Output Failure13
  • Output Problem12
  • Connection Problem3
  • Failure to Power Up3
  • Use of Device Problem3
  • Appropriate Device Problem Term/Code Not Available2
  • Defective Device2
  • Mechanical Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions37
  • No Known Impact Or Consequence To Patient4
  • Thrombosis/Thrombus1

Recent adverse event reports

The 25 most recent of 42 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AN ANGIODYNAMICS TERRITORY MANAGER REPORTED AN ISSUE WITH A VENACURE 1470 PRO LASER. THE END USER REPORTED THAT THEY RECENTLY HAD SEVERAL PATIENT HAVE EHITS {ENDOTHERMAL HEAT-INDUCED THROMBOSIS} DURING THEIR 1 WEEK FOLLOW UPS POST EVLT.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 19508093 ↗
  • MalfunctionAug 22, 2022Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATO WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15266221 ↗
  • MalfunctionAug 19, 2022No Display/Image

    DEVICE DISPALY NOT WORKING DURING TREATMENT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15256872 ↗
  • MalfunctionAug 19, 2022Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15256871 ↗
  • MalfunctionAug 9, 2022Connection Problem; Connection Problem

    THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15189935 ↗
  • MalfunctionJul 15, 2022Therapeutic or Diagnostic Output Failure

    THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15025448 ↗
  • MalfunctionJul 15, 2022Therapeutic or Diagnostic Output Failure

    THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15025431 ↗
  • MalfunctionJun 16, 2022Failure to Power Up; Failure to Power Up

    THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14711271 ↗
  • MalfunctionJun 14, 2022Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14684253 ↗
  • MalfunctionJun 14, 2022Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14684249 ↗
  • MalfunctionMay 23, 2022Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14468714 ↗
  • MalfunctionMay 12, 2022Defective Device

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14367715 ↗
  • MalfunctionApr 7, 2022Defective Device

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14039691 ↗
  • MalfunctionFeb 17, 2022Break; Break

    THE LASER HAD A FAILURE THAT DID NOT ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13547488 ↗
  • MalfunctionDec 21, 2021Mechanical Problem; Mechanical Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13050770 ↗
  • MalfunctionDec 16, 2021Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13019125 ↗
  • MalfunctionNov 16, 2021Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12818363 ↗
  • MalfunctionNov 11, 2021Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12795617 ↗
  • MalfunctionNov 11, 2021Therapeutic or Diagnostic Output Failure

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12792225 ↗
  • MalfunctionOct 28, 2021Defective Component

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12710455 ↗
  • MalfunctionOct 1, 2021Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12559937 ↗
  • MalfunctionSep 16, 2021Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12480937 ↗
  • MalfunctionJul 31, 2021Connection Problem; Connection Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12257185 ↗
  • MalfunctionJun 25, 2021Failure to Power Up; Failure to Power Up

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12068519 ↗
  • MalfunctionJun 23, 2021Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12050834 ↗

Other Powered Laser Surgical Instrument devices

All Powered Laser Surgical Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page