Skip to content

SLT II M094E1Recalls & reported problems

El.En. SPA · The Cynosure SLT II workstation has a solid state Nd:YAG laser capable of emitting laser radiation in three wavelenghts and includes a wide range of different delivery system, handpieces, fibers, cannulas and other accessories. · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
2
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08057017760009), received between Nov 18, 2016 and Feb 8, 2019 — classified as 2 deaths, 1 malfunction.

The most frequently reported specific device problem is “Output Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Output Problem1

Patient problems

  • Death2
  • Patient Problem/Medical Problem2
  • No Patient Involvement1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • DeathFeb 8, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON JANUARY 21ST, 2019, EL.EN. ELECTRONIC ENGINEERING (B)(4) BECAME AWARE OF AN ADVERSE EVENT, REPORTED BY US IMPORTER, (B)(4), THAT RECEIVED A COMMUNICATION FROM THE BROTHER OF THE PATIENT, CONCERNING THE DEATH OF THE PATIENT 4 DAYS AFTER A TREATMENT WITH SLT II LASER MEDICAL DEVICE. THE PATIENT'S BROTHER REPORTED TO THE US IMPORTER, ON JANUARY THE 17TH, 2019, THAT HER SISTER HAS DIED DUE TO BLEEDING FOLLOWING A TREATMENT WITH THE SLT II LASER MEDICAL DEVICE ON THE THIGHS AREA. WE, THE MANUFACTURER OF THE DEVICE, REQUESTED THE COOPERATION OF OUR US IMPORTER (B)(4) LOCATED IN (B)(4) US, FOR THE INVESTIGATION ON THIS CASE. (B)(4), AT THE TIME OF THE INITIAL COMMUNICATION TO US ALREADY TOOK CHARGE OF THE COMPLAINT AND IMMEDIATELY STARTED ITS OWN INVESTIGATION. (B)(4) ALSO REPRESENTS US DISTRIBUTOR AND SERVICE CENTER FOR EL.EN. ELECTRONIC ENGINEERING (B)(4) MEDICAL DEVICES. (B)(4) IDENTIFIED THE ACTUAL DEVICE INVOLVED IN THE EVENT AS AN SLT II LASER MEDICAL DEVICE SERIAL NUMBER #(B)(4), REF #M094E1 MANUFACTURED BY EL.EN. ELECTRONIC ENGINEERING (B)(4) COMPANY PLACED IN (B)(4). THE ACTUAL INVOLVED MEDICAL DEVICE IS INSTALLED AT (B)(6). (B)(4)'S CLINICAL STAFF PERFORMED THE INVESTIGATION

    Patient problems: Death; Patient Problem/Medical Problem

    MAUDE report 8319889 ↗
  • DeathFeb 6, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A PATIENT HAD DIED FOLLOWING A LASER PROCEDURE WITH THE SLT II DEVICE.

    Patient problems: Death; Patient Problem/Medical Problem

    MAUDE report 8313178 ↗
  • MalfunctionNov 18, 2016Output Problem; Output Problem

    US IMPORTER'S (B)(4) NOTICED US ABOUT A MALFUNCTION THEY RECENTLY BECAME AWARE OF. INVOLVED DEVICE IS CYNOSURE CELLULAZE - SLT II MODEL NUMBER M094A1, S/N (B)(4), MANUFACTURED BY (B)(4). THE ACTUAL DATE OF THE EVENT IS UNKNOWN (BEST ESTIMATE DATE OF THE EVENT (B)(6) 2016) AND EL.EN. SPA BECAME AWARE OF THE EVENT ON OCTOBER 20TH, 2016. EVENT TOOK PLACE IN THE US TERRITORY AT (B)(6). THE USER WAS COMPLAINING THAT THE DEVICE WAS TAKING MORE TIME THAN USUAL TO REACH THE TARGET TEMPERATURE FOR 'PRECISION' LASER TREATMENT OF FACE AND NECK WITH 1440 NM WAVELENGHT TAKING MORE JOULE TO REACH IT. US IMPORTER'S AUTHORIZED SERVICE TECHNICIAN VISITED THE CUSTOMER AND INFORMED US THAT THE SLT II DEVICE, DURING A SERVICE CHECK, FAILS TO MEET SPECIFICATIONS (LOW POWER ENERGY OUTPUT BELOW THE 20% OF TOLERANCE). NO PATIENT INVOLVED AND NO INJURY TO ANY PEOPLE. THIS MALFUNCTION COULD LEAD TO AN UNDER-TREATMENT BUT CANNOT, IN ANY TIME, LEAD TO A SERIOUS INJURY OR DEATH. THAT SAID, ACCORDING TO 'MEDICAL DEVICE REPORTING FOR MANUFACTURER - GUIDANCE FOR INDUSTRY AND FOOD AND DRUG ADMINISTRATION STAFF' ISSUED ON NOVEMBER 8TH, 2016, AT CHAPTER 2.14, THIS MALFUNCTION IS NOT REPORTABLE. WE, THE MANUFACTURER,

    Patient problems: No Patient Involvement

    MAUDE report 6112772 ↗

FDA clearance and classification

Cleared through 510(k) K123407 on Mar 29, 2013 (decision: Substantially Equivalent). Submitted by Cynosure, Inc. as “Cellulaze Laser”. The FDA review took 144 days. FDA record ↗

Cleared through 510(k) K102541 on Jan 26, 2012 (decision: Substantially Equivalent). Submitted by Cynosure, Inc. as “Cynosure Cellulaze Laser”. The FDA review took 510 days. FDA record ↗

Cleared through 510(k) K091537 on Sep 3, 2009 (decision: Substantially Equivalent). Submitted by Cynosure, Inc. as “Cynosure Smartlipo Multiwavelength Laser With 1440nm Wavelength”. The FDA review took 100 days. FDA record ↗

The FDA regulates product code OYW (laser, cellulite appearance) under 21 CFR 878.4810 as a Class II device. 21 CFR 878.4810 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the SLT II M094E1

Has the SLT II M094E1 been recalled?

No FDA recall record in our data names this exact device identifier (DI 08057017760009), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the SLT II M094E1?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Nov 18, 2016 and Feb 8, 2019. In those reports, the most common specific device problem is “Output Problem” (1); the most common patient problems are “Death” (2) and “Patient Problem/Medical Problem” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the SLT II M094E1?

2 reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the SLT II M094E1 FDA approved?

It was cleared through 510(k) K091537 on Sep 3, 2009; its record lists 3 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the SLT II M094E1, and is it still sold?

The FDA's device database lists El.En. SPA as the company responsible for it (brand SLT II, model M094E1). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page