Skip to content

DEKA MOTUS AZ+ M130A1Recalls & reported problems

El.En. SPA · The DEKA Motus AZ+ is a medical device equipped with Alexandrite 755nm laser and Nd:YAG 1064nm laser. The laser sources deliver the laser output through a lens coupled user replaceable optical fiber with a wide range of interchangeable, quick release laser handpieces with electronic spot recognition. · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08057017760566), received between Mar 8, 2024 and Mar 8, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Radiation Leak” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Radiation Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 8, 2024Radiation Leak; Radiation Leak

    ON FEBRUARY THE19TH, 2024, EL. EN. ELECTRONIC ENGINEERING SPA BECAME AWARE OF A MALFUNCTION, FROM CARTESSA'S (US IMPORTER) TECHNICIAN, THAT PERFORMED A SERVICE INTERVENTION ON THE MOTUS AZ+, INSTALLED AT WARNER INSTITUTE (ILLINOIS - USA) , IN WHICH THE SHUTTER BROKE AND CAUSED AN UNWANTED LASER EMISSION DURING THE LASER INITIAL CALIBRATION. THE ACTUAL DEVICE INVOLVED IS A MEDICAL DEVICE MOTUS AZ+ WHICH IS MARKETED IN THE US TERRITORIES WITH 510(K) K211821. IN THE COMMUNICATION RECEIVED BY THE US IMPORTER, CARTESSA, VIA A DEDICATED SERVICE REPORT CASE ID # (B)(4) (AND RELATIVE COMMUNICATION HELD WITH THE SERVICE DEPARTMENT) ARE REPORTED THE FOLLOWING INFORMATION: THE TECHNICIAN REPORTED TO HAVE PERFORMED A SERVICE ON THE DEVICE BECAUSE THE END USER WAS COMPLAINING OF AN INTERNAL TEST ERROR OF THE DEVICE. NO PATIENT OR OPERATOR SERIOUS INJURY HAS BEEN REPORTED. ONCE ARRIVED ON THE SITE THE TECHNICIAN NOTED THE INTERNAL TEST ERROR WHEN ATTEMPTED TO GO FORM THE 'STAND-BY' TO THE 'OPERATE/READY' STATE OF THE DEVICE. THE TECHNICIAN PROCEEDED TO CHECK, WHILE THE DEVICE PERFORMED THE INTERNAL CALIBRATION (AT THE TIME OF THE STATE PASSAGE), IF ANY EMISSION WAS PERFORMED AND FOUND OUT THAT R

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18866085 ↗

Other Powered Laser Surgical Instrument devices

All Powered Laser Surgical Instrument devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.