DEKA MOTUS AZ+ M130A1Recalls & reported problems
El.En. SPA · The DEKA Motus AZ+ is a medical device equipped with Alexandrite 755nm laser and Nd:YAG 1064nm laser. The laser sources deliver the laser output through a lens coupled user replaceable optical fiber with a wide range of interchangeable, quick release laser handpieces with electronic spot recognition. · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08057017760566), received between Mar 8, 2024 and Mar 8, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Radiation Leak” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Radiation Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 8, 2024Radiation Leak; Radiation Leak
ON FEBRUARY THE19TH, 2024, EL. EN. ELECTRONIC ENGINEERING SPA BECAME AWARE OF A MALFUNCTION, FROM CARTESSA'S (US IMPORTER) TECHNICIAN, THAT PERFORMED A SERVICE INTERVENTION ON THE MOTUS AZ+, INSTALLED AT WARNER INSTITUTE (ILLINOIS - USA) , IN WHICH THE SHUTTER BROKE AND CAUSED AN UNWANTED LASER EMISSION DURING THE LASER INITIAL CALIBRATION. THE ACTUAL DEVICE INVOLVED IS A MEDICAL DEVICE MOTUS AZ+ WHICH IS MARKETED IN THE US TERRITORIES WITH 510(K) K211821. IN THE COMMUNICATION RECEIVED BY THE US IMPORTER, CARTESSA, VIA A DEDICATED SERVICE REPORT CASE ID # (B)(4) (AND RELATIVE COMMUNICATION HELD WITH THE SERVICE DEPARTMENT) ARE REPORTED THE FOLLOWING INFORMATION: THE TECHNICIAN REPORTED TO HAVE PERFORMED A SERVICE ON THE DEVICE BECAUSE THE END USER WAS COMPLAINING OF AN INTERNAL TEST ERROR OF THE DEVICE. NO PATIENT OR OPERATOR SERIOUS INJURY HAS BEEN REPORTED. ONCE ARRIVED ON THE SITE THE TECHNICIAN NOTED THE INTERNAL TEST ERROR WHEN ATTEMPTED TO GO FORM THE 'STAND-BY' TO THE 'OPERATE/READY' STATE OF THE DEVICE. THE TECHNICIAN PROCEEDED TO CHECK, WHILE THE DEVICE PERFORMED THE INTERNAL CALIBRATION (AT THE TIME OF THE STATE PASSAGE), IF ANY EMISSION WAS PERFORMED AND FOUND OUT THAT R
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18866085 ↗
Other Powered Laser Surgical Instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.