Skip to content

DEKA LIPOAI M131A1Recalls & reported problems

El.En. SPA · The LipoAI is medical laser equipped with a short pulse 1444 nm Nd:YAG laser source. The device delivers the laser energy through a 600 um optical fiber. The fiber is protected by a small stainless-steel cannula (max. 1.4mm) and is inserted in the tissue through a small aperture in the skin. · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08057017760580), received between Aug 4, 2025 and Aug 4, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Use of Incorrect Control/Treatment Settings” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Incorrect Control/Treatment Settings1

Patient problems

  • Full thickness (Third Degree) Burn1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 4, 2025Use of Incorrect Control/Treatment Settings; Use of Incorrect Control/Treatment Settings

    IN DATE JULY THE 9TH 2025 WE RECEIVE A COMMUNICATION FROM THE US IMPORTER, (B)(6) RELATIVE TO AN ADVERSE EVENT IN WHICH A PATIENT DEVELOPED BURNS ON THE CHEST FOLLOWING A LASER LIPOLYSIS TREATMENT WITH THE DEVICE LIPO AI. THE ACTUAL DEVICE INVOLVED IN THE EVENT IS A LIPO AI, MANUFACTURED BY (B)(6) ELECTRONIC ENGINEERING SPA, MARKETED IN THE USA WITH 510(K) K212270. THE US IMPORTER, IN THE NARRATIVE AND INVESTIGATION PROVIDED WITH THE EVENT, REPORTED THE FOLLOWING: THE TREATMENT WAS PERFORMED IN DATE (B)(6) 2025. THE PARAMETERS USED FOR THE TREATMENT WERE 4887 JOULES DELIVERED ON EACH SIDE. 300MJ AND 40HZ (12 POWER AND LOW WERE LISTED ON THE OP NOTES). IT IS REPORTED THAT GLANDS HAVE BEEN REMOVED MANUALLY. PICTURES OF THE PATIENT WERE TAKEN THE DAY AFTER THE TREATMENT SHOWED SOME DARKENING UNDER THE TAPE. 4 DAYS AFTER TREATMENT THE PATIENT REMOVED THE TAPING THAT REVEALED FULL THICKNESS BURNS. IMMEDIATELY AFTER THE TREATMENT THE AREA WAS NOT FOUND WARM OR HOT. AS POST TREATMENT CARE THE PATIENT WAS INSTRUCTED TO NOT USE ANY GARMENT. WE, THE MANUFACTURER OF THE DEVICE, EVALUATED THE EVENT AS REPORTABLE UNDER 21 CFR PART 803 DUE TO THE FACT THAT THE LESION REPORTED BY THE PATIENT ARE

    Patient problems: Full thickness (Third Degree) Burn

    MAUDE report 22697367 ↗

Other Powered Laser Surgical Instrument devices

All Powered Laser Surgical Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page