V-STEADY 800039Recalls & reported problems
G21 Srl · Radiopaque Bone Cement for vertebral consolidation · Class 2 · Implant
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08058964723819), received between May 15, 2024 and May 15, 2024 — classified as 1 injury.
The most frequently reported specific device problem is “Migration” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Migration1
Patient problems
- Pulmonary Embolism1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMay 15, 2024Migration; Migration
ON (B)(6) 2021 FLUOROSCOPY GUIDED KYPHOPLASTY AT THE T3/T4 USING V-STEADY BONE CEMENT TO STABILIZE VERTEBRAL BODIES OCCURRED ON PATIENT. SURGEON STATES "RADIOGRAPHS WERE OBTAINED THAT SHOWED GOOD FILLING WITH NO POSTERIOR EXTRAVASATION INTO THE SPINAL CANAL" AND "FINAL RADIOGRAPHS WERE OBTAINED THAT DEMONSTRATED APPROPRIATE FILLING PATTERNS." THE FOLLOWING DAY THE PATIENT RETURNED TO THE ER WITH PAINFUL BREATHING UPON INHALATION, AND EXPERIENCED PAIN IN HER THORACIC AREA WITH COUGHING, ANY MOVEMENT, AND BREATHING. PATIENT'S COMPLICATIONS PERSISTED AND ON (B)(6) 2021 A CT WITH CONTRAST SHOWED EXTRAVASATION OF METHYL METHACRYLATE SHOWING EVIDENCE THAT CEMENT WAS PROPAGATING FROM THE T7 VERTEBRAL BODY INTO A PERIVERTEBRAL DRAINING VEIN AND INTO/AROUND/NEAR THE HEART. AN ECHO (TTE X 4) WAS PERFORMED WHICH CONFIRMED THAT THERE WAS CEMENT IN THE VEINS TO PATIENT'S HEART. PHYSICIAN STATED THE CEMENT LEAKED INTO THE AZYGOS VEIN INTO THE MAIN ATRIUM IN THE HEART. THE CT SCAN REVEALED: "1. LONG LINEAR HYPERDENSE MATERIAL EXTENDING FROM THE T3 AND AND T4 VERTEBRAL BODIES (STATUS POST CEMENT AUGMENTATION), THROUGH THE RIGHT AZYGOS, RIGHT ATRIUM, AND RIGHT VENTRICLE. THIS IS FAVORED TO ALL REPR
Patient problems: Pulmonary Embolism
MAUDE report 19323097 ↗
Other Cement, Bone, Vertebroplasty devices
- CONFIDENCE SPINAL CEMENT SYSTEM 283910000Medos International Sàrl · CEMENT, BONE, VERTEBROPLASTY5 deaths163 reports
- CONFIDENCE SPINAL CEMENT SYSTEM 283913000Medos International Sàrl · CEMENT, BONE, VERTEBROPLASTY8 deaths146 reports
- KYPHON® Xpede™ Bone Cement and KYPHON® Mixer Pack CX01BMedtronic Sofamor Danek, Inc. · CEMENT, BONE, VERTEBROPLASTY3 deaths67 reports
- SYNFLATE 03.804.701SSynthes GmbH · CEMENT, BONE, VERTEBROPLASTY60 reports
- KYPHON® HV-R® Bone Cement C01AMedtronic Sofamor Danek, Inc. · CEMENT, BONE, VERTEBROPLASTY57 reports
- iVAS 0705315000Stryker Corporation · CEMENT, BONE, VERTEBROPLASTY1 recall3 reports
- Cortoss 2101-0005Stryker Corporation · CEMENT, BONE, VERTEBROPLASTY1 recall3 reports
- iVAS 0705310000Stryker Corporation · CEMENT, BONE, VERTEBROPLASTY1 recall1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.