Olympus Empower H35 EMP-H35Recalls & reported problems
Quanta System SPA · Distributed by Gyrus ACMI, L.P. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This is the Olympus Empower H35 surgical laser system, made by Quanta System. It has 42 adverse event reports on file, all describing malfunctions—mostly the laser failing to produce output or shutting down so it couldn't be used during procedures, with no reported harm to patients. There are no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 42
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 42 adverse event reports that list this exact device (DI 08059173390038), received between Mar 30, 2020 and Nov 28, 2022 — classified as 42 malfunctions.
The most frequently reported specific device problem is “Output Problem” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Output Problem14
- Therapeutic or Diagnostic Output Failure8
- Defective Device4
- Energy Output Problem4
- Electrical /Electronic Property Problem3
- Appropriate Device Problem Term/Code Not Available2
- Defective Component2
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions39
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 25 most recent of 42 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 28, 2022Output Problem; Output Problem
THE DEVICE HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15868555 ↗MalfunctionNov 28, 2022Output Problem; Output Problem
THE DEVICE HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15868550 ↗MalfunctionNov 24, 2022Energy Output Problem; Energy Output Problem
THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT HAVE BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15860267 ↗MalfunctionSep 29, 2022Output Problem; Output Problem
THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15507656 ↗MalfunctionAug 24, 2022Output Problem; Output Problem
THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15281109 ↗MalfunctionJul 22, 2022Output Problem; Output Problem
THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15082143 ↗MalfunctionJul 13, 2022Defective Component
THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14998179 ↗MalfunctionJun 21, 2022Display or Visual Feedback Problem; Display or Visual Feedback Problem
THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14753698 ↗MalfunctionJun 21, 2022Output Problem; Output Problem
THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14753693 ↗MalfunctionJun 21, 2022Output Problem; Output Problem
THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14753684 ↗MalfunctionMay 27, 2022Defective Device
THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14520639 ↗MalfunctionApr 21, 2022Defective Device
CUSTOMER STATED THAT THE LASER DOESN'T RECOGNIZE OPTICAL FIBERS BUT, AFTER DEVICE TESTING, THE COMPLAINT WAS NOT CONFIRMED AND THE LASER WORKED PROPERLY. AS THE LITHOTRIPSY PROCEDURE LASTED ABOUT THREE HOURS, IT IS POSSIBLE THAT TWO FIBERS EXPIRED MAXIMUM DELIVERABLE ENERGY HAS BEEN REACHED. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14166133 ↗MalfunctionApr 14, 2022Use of Device Problem; Use of Device Problem
THE PROBLEM IS DUE TO A USE ERROR: THE DOCTOR PRESSED THE PEDAL IN STANDBY MODE. OBVIOUSLY NO LASER EMISSION IS POSSIBLE UNTIL SWITCHING FROM STANDBY TO READY MODE. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14110405 ↗MalfunctionApr 12, 2022Output Problem; Output Problem
THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14085523 ↗MalfunctionApr 7, 2022Defective Device
THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14039690 ↗MalfunctionMar 29, 2022Output Problem; Output Problem
THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13939410 ↗MalfunctionMar 24, 2022Energy Output Problem; Energy Output Problem
THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13890131 ↗MalfunctionMar 23, 2022Output Problem; Output Problem
THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13857163 ↗MalfunctionMar 17, 2022Output Problem; Output Problem
THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13800563 ↗MalfunctionFeb 16, 2022Break; Break
THE LASER HAD A FAILURE THAT DID NOT ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13542992 ↗MalfunctionFeb 4, 2022Energy Output Problem; Energy Output Problem
THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13451399 ↗MalfunctionJan 14, 2022Energy Output Problem; Energy Output Problem
THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13257142 ↗MalfunctionJan 14, 2022Therapeutic or Diagnostic Output Failure
THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13257137 ↗MalfunctionDec 17, 2021Defective Component
THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13024018 ↗MalfunctionDec 9, 2021Defective Device
THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12968707 ↗
Other Powered Laser Surgical Instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page