Olympus Empower H100 EMP-H100Recalls & reported problems

Quanta System SPA · Distributed by Gyrus ACMI, L.P. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus Empower H100 is a powered laser surgical instrument used during operations. The 56 reports mostly describe the laser stopping or failing to deliver output mid-procedure, sometimes with leaking fluid, which in one case postponed surgery. No recalls name this device, and reports are flat: none in the last 12 months or the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
56
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 56 adverse event reports that list this exact device (DI 08059173390052), received between Jan 17, 2020 and Mar 21, 2024 — classified as 56 malfunctions.

The most frequently reported specific device problem is “Therapeutic or Diagnostic Output Failure” (18 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Therapeutic or Diagnostic Output Failure18
  • Output Problem14
  • Appropriate Device Problem Term/Code Not Available7
  • Fluid/Blood Leak5
  • Electrical /Electronic Property Problem3
  • Calibration Problem2
  • Defective Device2
  • Energy Output Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions48
  • No Known Impact Or Consequence To Patient10

Recent adverse event reports

The 25 most recent of 56 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 21, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    DEVICE STOPPED WORKING IN THE MIDDLE OF SURGICAL PROCEDURE WITH PATIENT UNDER ANAESTHESIA WITH CONSEQUENT TREATMENT POSTPONEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18947912 ↗
  • MalfunctionOct 21, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15649158 ↗
  • MalfunctionJul 13, 2022Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14998175 ↗
  • MalfunctionJun 14, 2022Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14684245 ↗
  • MalfunctionMay 27, 2022Energy Output Problem; Energy Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14521018 ↗
  • MalfunctionMay 27, 2022Defective Device

    THE LASER SYSTEM HAD A FAILURE THAT DIDN'T ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT/OPERATOR WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14521015 ↗
  • MalfunctionApr 5, 2022Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14014955 ↗
  • MalfunctionMar 23, 2022Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13857110 ↗
  • MalfunctionMar 8, 2022Therapeutic or Diagnostic Output Failure

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13692316 ↗
  • MalfunctionFeb 10, 2022Therapeutic or Diagnostic Output Failure

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13505536 ↗
  • MalfunctionFeb 4, 2022Energy Output Problem; Energy Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13451386 ↗
  • MalfunctionJan 14, 2022Therapeutic or Diagnostic Output Failure

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13257145 ↗
  • MalfunctionJan 11, 2022Fluid/Blood Leak; Fluid/Blood Leak

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13227590 ↗
  • MalfunctionJan 5, 2022Fluid/Blood Leak; Fluid/Blood Leak

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY.NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13167971 ↗
  • MalfunctionDec 17, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13024109 ↗
  • MalfunctionNov 26, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12877221 ↗
  • MalfunctionSep 24, 2021Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12525379 ↗
  • MalfunctionSep 24, 2021Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12524837 ↗
  • MalfunctionSep 5, 2021Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12426510 ↗
  • MalfunctionSep 1, 2021Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12406370 ↗
  • MalfunctionAug 29, 2021Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12381412 ↗
  • MalfunctionAug 12, 2021Output Problem; Output Problem

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12311705 ↗
  • MalfunctionAug 10, 2021Failure to Power Up; Failure to Power Up

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12299237 ↗
  • MalfunctionJul 22, 2021Complete Loss of Power; Complete Loss of Power

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12213118 ↗
  • MalfunctionJul 15, 2021Failure to Power Up; Failure to Power Up

    THE LASER SYSTEM HAD A FAILURE THAT DID NOT ALLOW TO USE IT PROPERLY. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12177912 ↗

Other Powered Laser Surgical Instrument devices

All Powered Laser Surgical Instrument devices →