GEM Premier 5000 PAK BG/ISE/COOX 600 TEST 00055360008Recalls & reported problems
Instrumentation Laboratory Company · GEM Premier 5000 PAK pH, pCO2, pO2, Na+, K+, Cl-, Ca++, Hct, tHb, O2Hb, COHb, MetHb, HHb, sO2 - 600 Test 21 Days · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
24 other recalls cover devices cleared under the same FDA submission (K160225, K160415, K173403, K203790); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jan 20, 2026Class IIOpen, Classified
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Z-1525-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K160225, K160415, K173403, K203790. One submission often covers many models, so check the product and lot details in each record.
- Z-2900-2026 · May 21, 2026 · Open, ClassifiedGEM Premier 5000 PAK Part Numbers: 00055360010 00055360011 00055407504 00055415010 00055430010 00055430011 00055445010 00055445011 00055360010.A
- Z-2823-2026 · May 8, 2026 · Open, ClassifiedBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Product Description: The GEM Premier 5000 system provides fast, accurate, quantitative
- Z-2826-2026 · May 8, 2026 · Open, ClassifiedBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Product Description: The GEM Premier 5000 system provides fast, accurate, quantitative
- Z-2824-2026 · May 8, 2026 · Open, ClassifiedBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Product Description: The GEM Premier 5000 system provides fast, accurate, quantitative
- Z-2821-2026 · May 8, 2026 · Open, ClassifiedBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Intelligent Quality Management 2 (iQM2) is used as the quality control and assessment
- Z-2825-2026 · May 8, 2026 · Open, ClassifiedBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Product Description: The GEM Premier 5000 system provides fast, accurate, quantitative
- Z-1543-2026 · Jan 20, 2026 · Open, ClassifiedGEM Premier 5000; Part No. 00055445010 & 00055445011.
- Z-1531-2026 · Jan 20, 2026 · Open, ClassifiedGEM Premier 5000 PAK; Part No. 00055407511.
- Z-1528-2026 · Jan 20, 2026 · Open, ClassifiedGEM Premier 5000 PAK; Part No. 00055407504.
- Z-1524-2026 · Jan 20, 2026 · Open, ClassifiedGEM Premier 5000 PAK, Part No. 00055360004.
- Z-1534-2026 · Jan 20, 2026 · Open, ClassifiedGEM Premier 5000; Part No. 00055415008.
- Z-1541-2026 · Jan 20, 2026 · Open, ClassifiedGEM Premier 5000; Part No. 00055445004.
- Z-1538-2026 · Jan 20, 2026 · Open, ClassifiedGEM Premier 5000; Part No. 00055430008.
- Z-1542-2026 · Jan 20, 2026 · Open, ClassifiedGEM Premier 5000; Part No. 00055445008.
- Z-1537-2026 · Jan 20, 2026 · Open, ClassifiedGEM Premier 5000; Part No. 00055430004.
…and 9 more.
Other ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH devices
- GEM Premier 5000 00024019255Instrumentation Laboratory Company · Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph5 recalls5 deaths7 reports
- ABL800 FLEX 393-800Radiometer Medical ApS · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls81 reports
- ABL90 FLEX Analyzer 393-090Radiometer Medical ApS · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall112 reports
- i-STAT 03P88-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH105 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 300 TEST 00055430010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls3 deaths3 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 450 TEST 00055445010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls2 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 150 TEST 00055415010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX/TBILI 450 TEST 00055445011Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page