Skip to content

GEM Premier ChemSTAT PAK Na+,K+,Ca++,Cl-, Glu, Crea, BUN, tCO2, Hct, Lac,pH,pCO2 00044307510Recalls & reported problems

Instrumentation Laboratory Company · GEM Premier ChemSTAT PAK Na+, K+, Ca++, Cl-, Glu, Crea, BUN, tCO2, Hct, Lac, pH, pCO2 75 test PAK · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08426951070770), received between Mar 12, 2025 and Mar 12, 2025 — classified as 2 malfunctions.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Apparent Adverse Event2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 12, 2025No Apparent Adverse Event; No Apparent Adverse Event

    A CUSTOMER REPORTED THAT CREATININE RESULTS FROM A GEM PREMIER CHEMSTAT PAK (CARTRIDGE) WERE HIGHER THAN EXPECTED WHEN COMPARED TO RESULTS FROM THEIR LABORATORY ANALYZER (ROCHE COBAS). NO PATIENT IMPACT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21589410 ↗
  • MalfunctionMar 12, 2025No Apparent Adverse Event; No Apparent Adverse Event

    A CUSTOMER REPORTED THAT CREATININE RESULTS FROM A GEM PREMIER CHEMSTAT PAK (CARTRIDGE) WERE ERRONEOUSLY HIGH. PATIENTS WERE SENT FOR ADDITIONAL TESTING IN THE EMERGENCY DEPARTMENT WITH SAMPLES SENT TO THE LABORATORY (BECKMAN AU CREATININE JAFFE ASSAY), AND THE RESULTS WERE WITHIN NORMAL RANGE. NO PATIENT IMPACT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21587778 ↗

Other Electrode, Ion Specific, Sodium devices