The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08426951070794), received between Mar 12, 2025 and Nov 7, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device Ingredient or Reagent Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Ingredient or Reagent Problem1
  • Incorrect Measurement1
  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 7, 2025Incorrect Measurement; Device Ingredient or Reagent Problem; Incorrect Measurement; Device Ingredient or Reagent Problem

    POOR PERFORMANCE FOR CREATININE ON CHEMSTAT ON CHEMSTAT SYSTEM WAS OBSERVED DURING A RECENT EQA DISTRIBUTION (B)(4). IN ADDITION, PATIENT RESULT DISCREPANCIES WERE OBSERVED ON CARTRIDGE DETAILS. THE SAME CARTRIDGE WAS EXHIBITING MULTIPLE "SOLUTION DETECT ERROR DURING SAMPLE" AND REQUIRED REPLACEMENT ON 10/08/2025 DUE TO A "REFERENCE VOLTAGE ERROR". CUSTOMER REPORTED THAT THE NEW CARTRIDGE (BATCH 20908C, SERIAL NUMBER (B)(6) IS PERFORMING MORE IN LINE WITH THEIR CREATININE EXPECTATIONS. NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23500580 ↗
  • MalfunctionMar 12, 2025No Apparent Adverse Event; No Apparent Adverse Event

    A CUSTOMER REPORTED THAT CREATININE RESULTS FROM A GEM PREMIER CHEMSTAT PAK (CARTRIDGE) WERE DRIFTING HIGHER OUTSIDE THE REFERENCE RANGE FOR CREATININE: 0.80-1.40 MG/DL. SUBSEQUENTLY, THE CUSTOMER CONDUCTED A CORRELATION STUDY, AND RESULTS FROM THE GEM PREMIER CHEMSTAT PAK WERE 0.3-0.5 MG/DL HIGHER THAN THOSE OBTAINED USING THEIR CHEMISTRY METHOD (JAFFE CREATININE ASSAY). NO PATIENT IMPACT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions