TREO Abdominal Stent-Graft System 28-L2-17-100U (UDI 08435761052081)Recalls & reported problems
Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant
All models of TREO Abdominal Stent-Graft System (73)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08435761052081), received between Mar 19, 2026 and Apr 23, 2026 — classified as 2 injuries.
The most frequently reported specific device problem is “Device Endoleak” (2 reports).
1 other recall cover devices cleared under the same FDA submission (P190015); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Endoleak2
Patient problems
- Clinical Endoleak2
- Thrombosis/Thrombus1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 23, 2026Device Endoleak
"I WAS MADE AWARE LAST NIGHT OF AN EXPLANT OF A TREO DEVICE, PHYSICIAN STATED THERE WAS TYPE 3 ENDOLEAK AND THE DECISION WAS MADE TO EXPLANT THE DEVICE. NO FURTHER DETAILS WERE PROVIDED. ADDITIONAL INFORMATION RECEIVED ON 04/03/2026 FROM THE TERUMO AORTIC SITE REPRESENTATIVE: AT THE TIME OF IMPLANT BILATERAL RENAL ACCESSORIES WERE EMBOLIZED. POST IMPLANT ANGIOGRAM REVEALED SUCCESSFUL EXCLUSION OF ANEURYSM WITHOUT EVIDENCE OF ENDOLEAK. A CT PERFORMED ON (B)(6) 2021 SHOWED A POSSIBLE TYPE 2 ENDOLEAK "LIKELY DUE TO A NON-EMBOLIZED ACCESSORY LEFT RENAL", SAC WAS STABLE. HE HAD MULTIPLE INTERVENTIONS FOR A TYPE 2 ENDOLEAK WHICH LED TO A LAPAROSCOPIC LIGATION OF IMA AND MULTIPLE LUMBARS ON (B)(6) 2024. HE HAD A CV MRA ON (B)(6) 2025 THAT SHOWED SAC 7.2CM, THE REPORT ALSO STATED "RIGHT COMMON ILIAC- DISTAL SEGMENT IS ECTATIC, THE ILIAC LIMB OF AORTOILIAC STENT GRAFT EXTENDS INTO THE COMMON ILIAC ARTERY. INCOMPLETE APPOSITION OF THE DISTAL END OF THE STENT GRAFT." EXPLANT PERFORMED (B)(6) 2026. ON THE OP REPORT IT STATES 'SIGNIFICANT BLEEDING WAS NOTED FROM A DEFECT IN THE GRAFT FROM THE POSTERIOR MAIN BODY'" PATIENT OUTCOME: "NO INFO PROVIDED ON PATIENT. ADDITIONAL INFORMATION RECEIVED ON
Patient problems: Clinical Endoleak
MAUDE report 24967239 ↗InjuryMar 19, 2026Device Endoleak
"PREVIOUSLY REPORTED AS COMPLAINT (B)(4). BILATERAL ILIAC ARTERY GRAFT THROMBUS DISCOVERED DURING OFFICE VISIT ON (B)(6) 2023. PATIENT DEVELOPED THROMBUS ON AND ADJACENT TO STENT GRAFT WITH NO EVIDENCE OF DISTAL EMBOLUS. PATIENT WAS SUBSEQUENTLY PLACED ON ELIQUIS (5MG, 2X PER DAY), ASPIRIN (81MG PER DAY), AND ATORVASTATIN (40MG PER DAY) WITHOUT COMPLICATIONS. TPA-004-008 RETURNED FOR 1 YEAR FOLLOW-UP VISIT ON (B)(6) 2023 IN WHICH THE CTA REPORT FOUND "STABLE ANEURYSMAL SAX SIZE WITH PERSISTENT TYPE II ENDOLEAK. NEW NON-OCCLUSIVE THROMBUS IN THE PROXIMAL ASPECT OF THE LEFT COMMON ILIAC LIMB OF THE STENT-GRAFT. PER INVESTIGATOR, "PATIENT STILL HAS THROMBUS ON HIS GRAFT AND APPARENTLY IT NOW IS INVOLVING A SLIGHTLY DIFFERENT AREA THAT IT WAS BEFORE. IT IS UNCLEAR TO ME WHY THIS IS OCCURRING, AS THIS IS SOMETHING WE DO NOT GENERALLY SEE UNLESS THERE IS A LARGE DIAMETER CHANGE BETWEEN THE GRAFT COMPONENT, IT IS NOT IN THIS CASE." ADDENDUM FOR ADDITIONAL COMPLAINT 22OCT2024: THE CORE LAB SENT A NOTIFICATION ON 17OCT2024 OF SIGNIFICANT FINDING AFTER REVIEW OF SUBJECT TPA-004-008'S 2-YEAR CT IMAGING DATED (B)(6) 2024 NOTING A >5MM INCREASE IN ANEURYSM SAC SIZE WHEN COMPARED TO THE 30-DAY C
Patient problems: Clinical Endoleak; Thrombosis/Thrombus
MAUDE report 24642815 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P190015. One submission often covers many models, so check the product and lot details in each record.
- Z-1689-2023 · May 3, 2023 · Open, ClassifiedThe TREO ABDOMINAL STENT-GRAFT SYSTEM (TREO) consists of four types of implants, specifically a Main Bifurcated Stent-Graft, a Leg Extension Stent-Graft, a Proximal Cuff Stent-Graf
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page