TREO Abdominal Stent-Graft System 28-L2-11-140U (UDI 08435761052197)Recalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of TREO Abdominal Stent-Graft System (73)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08435761052197), received between Apr 2, 2025 and Apr 2, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Device Endoleak” (1 report).

1 other recall cover devices cleared under the same FDA submission (P190015); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Endoleak1
  • Device Slipped1

Patient problems

  • Clinical Endoleak1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 2, 2025Device Slipped; Device Endoleak; Device Slipped; Device Endoleak

    "SUBJECT (B)(6) HAD THEIR 3-YEAR FOLLOW-UP ON (B)(6) 2025 WITHOUT ANY COMPLAINTS. THE CT DATED (B)(6) 2025 NOTED A PERSISTENT SUSPECTED TYPE II ENDOLEAK, CONFIRMED AS SUSPECTED BY THE INVESTIGATOR INITIALLY, PROMPTING AN ANGIO WITH POSSIBLE REINTERVENTION. THE MAX DIAMETER OF LESION WAS CONFIRMED BY THE INVESTIGATOR AS 55MM, WHICH IS THE SAME AS THE PRIOR 2-YEAR FOLLOW-UP AND <5MM INCREASE COMPARED TO 30-DAY FOLLOW-UP. THE INVESTIGATOR ADDITIONALLY NOTED A 10MM MIGRATION OF THE DEVICE, HOWEVER IT WAS STILL DEEMED PATENT. THE SUBJECT WAS ADMITTED ON (B)(6) 2025 AND UNDERWENT ANGIO WITH REINTERVENTION DUE TO WHAT WAS DETERMINED TO BE A NEW TYPE IA ENDOLEAK BY THE INVESTIGATOR. A SMALL ENDOLEAK ON THE RIGHT POSTERIOR ASPECT OF THE GRAFT WAS NOTED PROMPTING PLACEMENT OF A 30MM TERUMO AORTIC CUFF (LOT: 2402010416) PROXIMALLY. THERE WERE NO COMPLICATIONS DURING THE PROCEDURE. THE EVENT WAS ASSESSED AS RELATED TO DEVICE AND INDEX PROCEDURE. THE INVESTIGATOR FURTHER SPECIFIED THERE WAS NOT A DEVICE DEFICIENCY NOR DID THE REINTERVENTION OCCUR DUE TO A DEVICE DEFICIENCY." PATIENT OUTCOME: "THE SUBJECT WAS DISCHARGED ON (B)(6)2025 IN STABLE CONDITION, BUT THE EVENT IS CURRENTLY ONGOING. ADDIT

    Patient problems: Clinical Endoleak

    MAUDE report 21756272 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P190015. One submission often covers many models, so check the product and lot details in each record.

  • Z-1689-2023 · May 3, 2023 · Open, Classified
    The TREO ABDOMINAL STENT-GRAFT SYSTEM (TREO) consists of four types of implants, specifically a Main Bifurcated Stent-Graft, a Leg Extension Stent-Graft, a Proximal Cuff Stent-Graf

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →