Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08435761052456), received between Dec 18, 2024 and Feb 26, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).

1 other recall cover devices cleared under the same FDA submission (P190015); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information2
  • Adverse Event Without Identified Device or Use Problem1
  • Lack of Effect1
  • Patient Device Interaction Problem1

Patient problems

  • Aneurysm expansion2
  • Insufficient Information1
  • Ischemia1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 26, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Patient Device Interaction Problem; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Patient Device Interaction Problem

    "THE CORE LAB SENT NOTIFICATION ON 03FEB2025 REGARDING SIGNIFICANT FINDING FOR SUBJECT TPA-021-001'S 2-YEAR CT DATED (B)(6) 2024 REFLECTING A >5MM INCREASE IN ANEURYSM SAC SIZE COMPARED TO THE 30-DAY DATED 07NOV2022. ON (B)(6) 2024, THE SUBJECT PRESENTED FOR THEIR 2-YEAR FOLLOW-UP WITHOUT COMPLAINTS OF CLAUDICATION AND AN OVERALL STABLE VASCULAR EXAM. HIS CT REFLECTED A ~7MM INCREASE IN SAC SIZE WITH A MAX DIAMETER OF 65MM COMPARED TO THE 30-DAY CT WITH A MAX DIAMETER OF 58.5MM, AS CONFIRMED BY THE INVESTIGATOR. THE COMMON AND EXTERNAL ILIAC ARTERIES ARE PATENT. THERE IS HIGH-GRADE STENOSIS OF THE RIGHT INTERNAL ILIAC ARTERY, AND THERE IS ALSO EVIDENCE OF RIGHT INTERNAL ILIAC ARTERY ANEURYSMAL EXPANSION TO [PER INVESTIGATOR] "ABOUT 30MM ON THE SAGITTAL VIEW. THIS HAS ALWAYS MEASURED 20-24MM IN THE PAST". THERE HAS BEEN NO EVIDENCE OF ENDOLEAK ON ANY FOLLOW-UP IMAGING (30-DAY, 1-YEAR, 2-YEAR) AFTER THEIR INDEX PROCEDURE. THE INVESTIGATOR ASSESSED THE EVENT AS POSSIBLY RELATED TO DEVICE. NOTE THAT AN EVENT INVOLVING THESE DEVICES WAS PREVIOUSLY REPORTED AND INVESTIGATED UNDER COMPLAINT NUMBER (B)(4). THE FOLLOWING ADDITIONAL INFORMATION WAS PROVIDED BY (B)(6) ON (B)(6) 2025. ON 25FEB

    Patient problems: Aneurysm expansion; Insufficient Information; Ischemia

    MAUDE report 21472763 ↗
  • InjuryDec 18, 2024Insufficient Device Problem Information; Lack of Effect; Insufficient Device Problem Information; Lack of Effect

    "THE CORE LAB SENT NOTIFICATION OF A SIGNIFICANT FINDING FOR SUBJECT (B)(6) ON 21NOV2024 REGARDING A >5MM INCREASE SAC ENLARGEMENT FOR 3-YEAR CT IMAGING DATED (B)(6) 2024 AS COMPARED TO THE 30-DAY IMAGING. THE CORE LAB'S ASSESSMENT REFLECTED A MAXIMUM ~7.7MM INCREASE. PENDING SITE DETAILS. 01" PATIENT OUTCOME: "PENDING SITE DETAILS."

    Patient problems: Aneurysm expansion

    MAUDE report 20969268 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P190015. One submission often covers many models, so check the product and lot details in each record.

  • Z-1689-2023 · May 3, 2023 · Open, Classified