DOSI-FUSER L25915-100D2-USARecalls & reported problems
Epic International(Thailand) Company Limited · Elastomeric Infusion Pump · Class 2
All models of DOSI-FUSER (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08436020763861), received between May 24, 2018 and May 28, 2018 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Insufficient Flow or Under Infusion” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Insufficient Flow or Under Infusion2
- Burst Container or Vessel1
- Failure to Infuse1
- Fluid/Blood Leak1
- Loss of or Failure to Bond1
- Partial Blockage1
Patient problems
- No Known Impact Or Consequence To Patient4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 28, 2018Burst Container or Vessel; Fluid/Blood Leak; Burst Container or Vessel; Fluid/Blood Leak
PATIENT STARTED 5-FU CONTINUOUS INFUSION AS PART OF FOLFOX CHEMOTHERAPY REGIMEN TO BE DELIVERED VIA DOSI-FUSER. (TOTAL VOLUME 86.3 ML; 100 ML/2 DAY/1.9 ML PER HOUR). PATIENT OBSERVED IN ONCOLOGY CLINIC WITH MALFUNCTIONED PUMP. LNNER BALLOON HAD BURST AT SOME POINT (PATIENT NOT ABLE TO DETERMINE WHEN) AND DRUG MOVING FREELY THROUGH CHAMBER, DRUG NOT DELIVERED TO PATIENT. NDC: 100 ML 2 DAY PUMP 1.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7547869 ↗MalfunctionMay 24, 2018Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion
PATIENT'S CHEMOTHERAPY ORDERED FOR 3750 MG TO INFUSE OVER 46 HOURS USING A DOSI-FUSER 100 AT 1.9 ML/HR. PATIENT WAS SEEN TO HAVE INFUSER DISCONTINUED AND IT WAS NOTED IT HAD NOT INFUSED COMPLETELY. THE PATIENT'S PORT WAS FLUSHED WITHOUT DIFFICULTY AND THE INFUSER WAS RECONNECTED. THE PATIENT RETURNED THE NEXT DAY AND THE INFUSER WAS RECONNECTED. THE PATIENT RETURNED THE NEXT DAY AND THE INFUSER REMAINED HALF FULL. THE PORT WAS FLUSHED AGAIN WITHOUT DIFFICULTY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7540766 ↗MalfunctionMay 24, 2018Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion
PATIENT'S DOSI-FUSER (100 ML AT 1.9ML/HR FOR 46 HOURS) DID NOT DELIVER PATIENT CHEMOTHERAPY PROPERTY. PER PHARMACIST THE PATIENT ONLY RECEIVED HALF OF THE INTENDED DOSE. PATIENT'S PORT LOCATED IN THE RIGHT UPPER CHEST FLUSHED WITH NORMAL SALINE WITHOUT DIFFICULTY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7540280 ↗MalfunctionMay 24, 2018Partial Blockage; Loss of or Failure to Bond; Failure to Infuse; Partial Blockage; Loss of or Failure to Bond; Failure to Infuse
PATIENT PRESENTED TO CLINIC BECAUSE HER PUMP WAS NO LONGER INFUSING MEDICATION. LABEL INDICATING VOLUME AND INFUSION RATE PROVIDED BY THE MANUFACTURER INSIDE THE PUMP CAPSULE WAS BENT AND WAS IMPEDING THE LEVEL INDICATOR. CONTACTED THE MFR AND THEY INDICATED THAT THE PUMPS WERE RECENTLY UPGRADED PUTTING SO THAT THE LABELS WERE NOW INSIDE RATHER THAN OUTSIDE THE PUMP. PUMP HAD TO BE DISCONNECTED AND MEDICATION HAD TO BE PUT IN A 24 HOUR PUMP.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7539912 ↗
Other Pump, Infusion, Elastomeric devices
- EASYPUMP 4540018-02B. Braun Medical Inc. · PUMP, INFUSION, ELASTOMERIC313 reports
- Halyard 13568Avanos Medical, Inc. · PUMP, INFUSION, ELASTOMERIC86 reports
- ON-Q* Pain Relief System with SELECT-A-FLOW 13472Avanos Medical, Inc. · Pump, infusion, elastomeric73 reports
- Halyard 13472Avanos Medical, Inc. · PUMP, INFUSION, ELASTOMERIC68 reports
- EASYPUMP 4540008-02B. Braun Medical Inc. · PUMP, INFUSION, ELASTOMERIC54 reports
- DOSI-FUSER L25915-600D1-USALeventon Sau · Pump, Infusion, Elastomeric1 recall0 reports
- DOSI-FUSER L25915-500D1-USALeventon Sau · Pump, Infusion, Elastomeric1 recall0 reports
- Halyard 13561Avanos Medical, Inc. · PUMP, INFUSION, ELASTOMERIC42 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page