Erytra 210401Recalls & reported problems
Diagnostic Grifols SA · The Erytra is a fully-automated high-throughput analyzer designed to automate in vitro immunohematologial test of human blood utilizing DG Gel 8 card technology, including Blood Grouping, Antigen Typing, Antibody Screening, Antibody Identification, Compatability Tests, and Direct Antiglobulin Tests. As a standalone analyzer or interfaced to the customer's Laboratory Information System (LIS), the Erytra automates test processing functions and data management requirements using DG Gel 8 cards and digital image processing. · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08437014205190), received between Jun 4, 2020 and Jun 4, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Device Displays Incorrect Message1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 4, 2020Device Displays Incorrect Message; Device Displays Incorrect Message
ON (B)(6) 2020, THE CUSTOMER (B)(6), REPORTED TO GRIFOLS A SAMPLE BARCODE READING ERROR OCCURRED DURING THE VALIDATION/CORRELATION OF A CROSSMATCH TEST. THE ERYTRA FULLY AUTOMATED IMMUNOHEMATOLOGY ANALYZER READ THE DONOR SAMPLE BARCODE AS (B)(6) INSTEAD OF THE EXPECTED NUMBER (B)(6). THE EVENT WAS DETECTED BY THE OPERATOR WHEN CHECKING THE BARCODES TO REPEAT THE CROSSMATCH TESTS MANUALLY. THE SAME BARCODE WAS LATER READ AGAIN WITH THE SAME INSTRUMENT AND WITH ANOTHER ERYTRA ANALYZER, AND THE BARCODE WAS READ CORRECTLY. THE CUSTOMER CONFIRMS THAT THIS WAS THE FIRST AND ONLY TIME THE DEVICE HAS MISREAD A BARCODE NUMBER. THE BARCODE WAS AN ISBT128 WITH CHECKSUM FROM A DONOR'S BAG FROM THE RED CROSS UNIT. NOTE THAT WHEN THE CUSTOMER IS WORKING IN ROUTINE CONDITIONS, THE INSTRUMENT IS INTERFACED WITH THE LIS. THE LIS WOULD DETECT THE DISCREPANCY BECAUSE THE WRONG CODE EITHER WOULD NOT EXIST OR WOULD NOT CORRESPOND TO A SAMPLE PREVIOUSLY SENT TO THE ANALYZER, PRECLUDING THE POSSIBILITY OF RELEASING THE ASSOCIATED TESTS RESULT TO AN INCORRECT SAMPLE IDENTIFICATION NUMBER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10117514 ↗
Other System, Test, Automated Blood Grouping And Antibody devices
- Galileo Echo 0087000Immucor, Inc. · System, Test, Automated Blood Grouping And Antibody242 reports
- ORTHO 6904577Ortho-Clinical Diagnostics, Inc. · SYSTEM, TEST, AUTOMATED BLOOD GROUPING AND ANTIBODY117 reports
- Galileo Neo 0064599Immucor, Inc. · System, Test, Automated Blood Grouping And Antibody65 reports
- TANGO infinity 850000010Bio-Rad Medical Diagnostics GmbH · System, Test, Automated Blood Grouping And Antibody1 death55 reports
- Echo Lumena 0086998Immucor, Inc. · System, Test, Automated Blood Grouping And Antibody49 reports
- IH-1000 001100Bio-Rad Medical Diagnostics GmbH · System, Test, Automated Blood Grouping And Antibody44 reports
- NEO Iris 0064597Immucor, Inc. · System, Test, Automated Blood Grouping And Antibody28 reports
- TANGO optimo 848900010Bio-Rad Medical Diagnostics GmbH · System, Test, Automated Blood Grouping And Antibody24 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.