Percuflex™ M001235600Recalls & reported problems
Boston Scientific Corporation · Ureteral Stent System with Dissolving Tip · Class 2
All models of Percuflex (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729031857), received between Mar 14, 2023 and Apr 6, 2023 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Detachment of Device or Device Component1
- Entrapment of Device1
- Material Split, Cut or Torn1
- Material Twisted/Bent1
- Migration1
- Positioning Failure1
- Positioning Problem1
Patient problems
- Foreign Body In Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 6, 2023Break; Positioning Problem; Break; Positioning Problem
IT WAS REPORTED VIA ATTACHED MEDWATCH (B)(4) THAT STENT WAS DIFFICULT TO POSITION AND SUTURE DETACHMENT OCCURRED WHICH REQUIRED ADDITIONAL INTERVENTION. A PERCUFLEX URETERAL STENTS WAS USED IN THE URETER. DURING THE PROCEDURE, THE NEPHROURETERAL CATHETER WAS SUCCESSFULLY REMOVED OVER A GUIDEWIRE AND THEN PLACED AN 11 FRENCH PEEL-AWAY SHEATH. A 10 FRENCH BY 26-CM DOUBLE-J STENT WAS THEN ADVANCED OVER THE GUIDEWIRE. AFTER DEPLOYMENT OF THE STENT, IT PROMPTLY MIGRATED DISTALLY WITH THE DEFORMED PROXIMAL RETENTION LOOP "KINKING" IN THE URETER APPROXIMATELY 5-7 CM FROM THE RENAL PELVIS. THE EXTERNAL RETENTION "SUTURE" WAS FOUND TO BE UNASSOCIATED/DETACHED FROM THE PROXIMAL ASPECT OF THE DOUBLE J STENT UPON APPLYING TENSION TO IT. EXTENSIVE ATTEMPTS TO REPOSITION THE STENT USING VASCULAR SNARES AND MULTIPLE CATHETER AND GUIDEWIRE COMBINATIONS WERE ATTEMPTED BUT WERE UNSUCCESSFUL. THE STENT COULD NOT BE APPROPRIATELY ENGAGED AT ITS PROXIMAL ASPECT. MULTIPLE ATTEMPTS WERE MADE AT PASSAGE OF A GUIDEWIRE INTO THE STENT, WITHOUT SUCCESS. THEREFORE, A GUIDEWIRE WAS ADVANCED INTO THE URINARY BLADDER AND A VASCULAR SNARE WAS USED TO ENGAGE THE DISTAL ASPECT OF THE DOUBLE-J STENT WITHIN THE URINA
Patient problems: Foreign Body In Patient
MAUDE report 16694189 ↗MalfunctionMar 14, 2023Positioning Failure; Entrapment of Device; Detachment of Device or Device Component; Material Twisted/Bent; Migration; Material Split, Cut or Torn
THE INTERVENTIONAL RADIOLOGIST SCHEDULED THE PATIENT FOR A PROCEDURE DURING WHICH HE INTENDED TO REMOVE THE PATIENT'S LEFT NEPHROURETERAL CATHETER AND REPLACE IT WITH A DOUBLE-J STENT. THE PHYSICIAN HAD SUCCESSFULLY REMOVED THE NEPHROURETERAL CATHETER OVER A GUIDEWIRE AND THEN PLACED AN 11 FRENCH PEEL-AWAY SHEATH. A 10 FRENCH BY 26-CM DOUBLE-J STENT WAS THEN ADVANCED OVER THE GUIDEWIRE. AFTER DEPLOYMENT OF THE STENT, IT PROMPTLY MIGRATED DISTALLY WITH THE DEFORMED PROXIMAL RETENTION LOOP "KINKING" IN THE URETER APPROXIMATELY 5-7 CM FROM THE RENAL PELVIS. THE EXTERNAL RETENTION "SUTURE" WAS FOUND TO BE UNASSOCIATED/DETACHED FROM THE PROXIMAL ASPECT OF THE DOUBLE J STENT UPON APPLYING TENSION TO IT. EXTENSIVE ATTEMPTS TO REPOSITION THE STENT USING VASCULAR SNARES AND MULTIPLE CATHETER AND GUIDEWIRE COMBINATIONS WERE ATTEMPTED BUT WERE UNSUCCESSFUL. THE STENT COULD NOT BE APPROPRIATELY ENGAGED AT ITS PROXIMAL ASPECT. MULTIPLE ATTEMPTS WERE MADE AT PASSAGE OF A GUIDEWIRE INTO THE STENT, WITHOUT SUCCESS. THEREFORE, A GUIDEWIRE WAS ADVANCED INTO THE URINARY BLADDER AND A VASCULAR SNARE WAS USED TO ENGAGE THE DISTAL ASPECT OF THE DOUBLE-J STENT WITHIN THE URINARY BLADDER. A SECOND 5 FRENC
Patient problems: Foreign Body In Patient
MAUDE report 16536613 ↗
Other STENT, URETERAL devices
- Bard® InLay Optima® Ureteral Stent 788426C. R. Bard, Inc. · STENT, URETERAL1 recall100 reports
- Bard® InLay Optima® Ureteral Stent 788626C. R. Bard, Inc. · STENT, URETERAL122 reports
- Bard® InLay® Ureteral Stent with HydroGlide™ Guidewire 777626C. R. Bard, Inc. · STENT, URETERAL118 reports
- Bard® InLay® Ureteral Stent with HydroGlide™ Guidewire 777426C. R. Bard, Inc. · STENT, URETERAL108 reports
- Bard® InLay® Ureteral Stent 778626C. R. Bard, Inc. · STENT, URETERAL100 reports
- Percuflex™ Plus M0061752530Boston Scientific Corporation · STENT, URETERAL1 death99 reports
- Contour VL™ M0061801560Boston Scientific Corporation · STENT, URETERAL1 recall45 reports
- Polaris™ Ultra M0061921330Boston Scientific Corporation · STENT, URETERAL71 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page