Percuflex™ Plus M0061752640Recalls & reported problems
Boston Scientific Corporation · Ureteral Stent · Class 2 · Implant
All models of Percuflex Plus (35)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 08714729041191), received between Dec 10, 2018 and Aug 8, 2025 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Adverse Event Without Identified Device or Use Problem1
- Deformation Due to Compressive Stress1
- Detachment of Device or Device Component1
- Difficult to Advance1
- Material Split, Cut or Torn1
- Packaging Problem1
Patient problems
- No Consequences Or Impact To Patient4
- No Clinical Signs, Symptoms or Conditions3
- Pain1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 8, 2025Break; Packaging Problem; Break; Packaging Problem
IT WAS REPORTED THAT A PERCUFLEX PLUS URETERAL STENT WAS USED IN A LASER LITHOTRIPSY PROCEDURE. DURING THE PROCEDURE, THE STENT WAS FRACTURED. IT WAS NOTED THAT A BEND IN THE PACKAGING HAVE CAUSED THE DAMAGE OF THE STENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER SAME DEVICE AND THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22746406 ↗MalfunctionJan 8, 2024Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS OPENED TO BE USED DURING A URETERAL CATHETERIZATION PROCEDURE IN THE URETER PERFORMED ON (B)(6) 2023. DURING PREPARATION, THE COIL OF THE STENT WAS FOUND FRACTURED. IT WAS NOTED THAT THE STENT WAS BROKEN INTO TWO PIECES ALONG THE SHAFT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER PERCUFLEX PLUS URETERAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18471401 ↗InjuryMay 20, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NOTE: THIS REPORT PERTAINS TO THE FIRST OF TWO DEVICES THAT WERE USED IN THE SAME PROCEDURE. IT WAS TO REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TWO PREVIOUSLY IMPLANTED PERCUFLEX PLUS URETERAL STENT WAS REMOVED FROM THE LEFT AND RIGHT URETER, PERFORMED ON (B)(6) 2022 AS PART OF THE U0652 DOUBLE-J REGISTRY CLINICAL STUDY. ON (B)(6) 2022, THE PERCUFLEX PLUS URETERAL STENT WAS SUCCESSFULLY PLACED FOR STONE MANAGEMENT IN THE LEFT AND RIGHT URETER, AND THE PATIENT WAS PRESCRIBED THE DISCHARGE MEDICATIONS ACETAMINOPHEN (TYLENOL) AND NARCOTIC ANALGESICS (OPIOIDS). NO ISSUES WERE NOTED WITH THE DEVICE. ON (B)(6) 2022, THE PATIENT HAD EXPERIENCED A LEFT AND RIGHT FLANK PAIN. THE PATIENT WAS GIVEN TYLENOL, IBUPROFEN, URINE SAMPLE, AND FLOMAX TO TREAT THE PAIN. DURING THE PLANNED STENT REMOVAL PERFORMED ON (B)(6) 2022, THE PERCUFLEX PLUS URETERAL STENT WAS SUCCESSFULLY REMOVED FROM THE PATIENT WITHOUT ANY DIFFICULTY. THERE WERE NO NEW DEVICE IMPLANTED. ON APRIL 19, 2022 THE EVENT WAS CONSIDERED RESOLVED.
Patient problems: Pain
MAUDE report 14458841 ↗MalfunctionNov 7, 2021Deformation Due to Compressive Stress; Difficult to Advance; Deformation Due to Compressive Stress; Difficult to Advance
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A PERCUTANEOUS NEPHROLITHOTOMY WITH ANTEGRADE STENT PLACEMENT PROCEDURE IN THE KIDNEY AND URETER, PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE AND INSIDE THE PATIENT, WHEN THE PHYSICIAN PLACED THE STENT OVER THE GUIDEWIRE, IT WAS OBSERVED THAT THE STENT WAS BUCKLED. THE PHYSICIAN TRIED TO REMOVE THE WIRE ONCE IDEAL PLACEMENT WAS ACHIEVED, THE WIRE WAS STUCK TO THE STENT AND THE GUIDEWIRE WAS UNABLE TO BE REMOVED. ANOTHER PERCULFEX PLUS URETERAL STENT WAS OPENED AND SUCCESSFULLY COMPLETED THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12759737 ↗MalfunctionOct 21, 2020Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN PERCUFLEX PLUS URETERAL STENT WAS OPENED TO BE USED DURING A PROCEDURE, PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, WHEN THE PHYSICIAN UNPACKED THE PACKAGE, IT WAS NOTICED THAT THE STENT WAS FRACTURED IN TWO PIECES. ANOTHER PERCUFLEX PLUS URETERAL STENT WAS OPENED AND SUCCESSFULLY COMPLETED THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10714777 ↗MalfunctionJul 8, 2020Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A URETEROSCOPY WITH STENT PLACEMENT PROCEDURE IN THE URINARY TRACT, KIDNEY TO BLADDER PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN THE STENT WAS INSERTED, THE PHYSICIAN NOTICED THAT THE BLADDER COIL WOULD NOT PROPERLY FORM A COIL. WHEN THE PHYSICIAN EXAMINED THE STENT VIA THE CAMERA LENS, A TEAR IN THE BLADDER COIL PORTION OF THE STENT WAS NOTICED. REPORTEDLY, THE ASSISTANT PULLED THE SUTURE AND INADVERTENTLY RIPPED THROUGH THE STENT, HOWEVER THE SUTURE REMAINED INTACT. THE STENT ITSELF WAS NOT SEPARATED INTO TWO PIECES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER PERCUFLEX PLUS URETERAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10244339 ↗MalfunctionMay 22, 2019Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS STENT WAS TO BE USED FOR A URETEROSCOPY PROCEDURE TO BE PERFORMED IN THE URETER ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING UNPACKING, THE STENT WAS REPORTED TO BE BROKEN. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH THE USE OF ANOTHER PERCUFLEX PLUS STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THE EVENT. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8633522 ↗MalfunctionDec 10, 2018Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS STENT WAS TO BE USED FOR A STENT PLACEMENT PROCEDURE PERFORMED IN THE URETER ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE SURGEON NOTED THAT THE DISTAL TIP OF THE STENT WAS DETACHED INSIDE OF THE PATIENT. THE STENT WAS REMOVED TOGETHER WITH THE SCOPE . THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE CONDITION OF THE PATIENT AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8147383 ↗
Other STENT, URETERAL devices
- Bard® InLay Optima® Ureteral Stent 788426C. R. Bard, Inc. · STENT, URETERAL1 recall100 reports
- Bard® InLay Optima® Ureteral Stent 788626C. R. Bard, Inc. · STENT, URETERAL122 reports
- Bard® InLay® Ureteral Stent with HydroGlide™ Guidewire 777626C. R. Bard, Inc. · STENT, URETERAL118 reports
- Bard® InLay® Ureteral Stent with HydroGlide™ Guidewire 777426C. R. Bard, Inc. · STENT, URETERAL108 reports
- Bard® InLay® Ureteral Stent 778626C. R. Bard, Inc. · STENT, URETERAL100 reports
- Percuflex™ Plus M0061752530Boston Scientific Corporation · STENT, URETERAL1 death99 reports
- Contour VL™ M0061801560Boston Scientific Corporation · STENT, URETERAL1 recall45 reports
- Polaris™ Ultra M0061921330Boston Scientific Corporation · STENT, URETERAL71 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page