Contour VL™ M006180155010Recalls & reported problems

Boston Scientific Corporation · Variable Length Ureteral Stent Set · Class 2 · Implant

All models of Contour VL (12)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729058427), received between Feb 18, 2022 and Jul 15, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Split, Cut or Torn4
  • Break2
  • Deformation Due to Compressive Stress2
  • Difficult to Advance2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 15, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT, DURING A RETROGRADE INTRARENAL SURGERY PROCEDURE, A CONTOUR VL STENT WAS USED. IT WAS FOUND THAT THE TIP OF THE BLADDER SIDE WAS TORN WHEN ADVANCING THE PUSHER FOR STENT PLACEMENT. ANOTHER OF THE SAME STENT WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE NOTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19744307 ↗
  • MalfunctionJul 12, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT, DURING RETROGRADE INTRARENAL SURGERY, A CONTOUR VL STENT WAS USED. IT WAS FOUND THAT THE TIP OF THE BLADDER SIDE WAS TORN WHEN ADVANCING THE PUSHER FOR STENT PLACEMENT. ANOTHER OF THE SAME STENT WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE NOTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19735419 ↗
  • MalfunctionDec 13, 2023Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CONTOUR VL URETERAL STENT WAS OPENED TO BE USED DURING A STONE LASER CASE PROCEDURE PERFORMED ON (B)(6) 2023. DURING PREPARATION, WHEN THE DEVICE WAS UNPACKED, THE STENT WAS FOUND BROKEN IN TWO PIECES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER CONTOUR VL URETERAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18320682 ↗
  • MalfunctionJul 8, 2022Deformation Due to Compressive Stress; Difficult to Advance; Adverse Event Without Identified Device or Use Problem; Material Split, Cut or Torn; Deformation Due to Compressive Stress; Difficult to Advance; Adverse Event Without Identified Device or Use Problem; Material Split, Cut or Torn

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CONTOUR VL URETERAL STENT WAS USED DURING A REMOVAL OF KIDNEY STONES PROCEDURE IN THE RIGHT KIDNEY PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, WHEN THE STENT WAS INTRODUCED AND PASSED THROUGH THE URETHRA AND ENTERED THE FORCED OF TWO CENTIMETERS INTO THE URETER, AND WHEN THE STENT WAS PUSHED, IT BECAME BUCKLED AND WOULD NOT ADVANCE INTO THE URETER AND A TORN WAS OBSERVED ON THE DISTAL PART. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER CONTOUR VL URETERAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14955455 ↗
  • MalfunctionJul 7, 2022Deformation Due to Compressive Stress; Difficult to Advance; Material Split, Cut or Torn; Deformation Due to Compressive Stress; Difficult to Advance; Material Split, Cut or Torn

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CONTOUR VL URETERAL STENT WAS USED DURING A REMOVAL OF KIDNEY STONES PROCEDURE IN THE RIGHT KIDNEY PERFORMED ON (B)(6), 2022. DURING THE PROCEDURE, WHEN THE STENT WAS INTRODUCED AND PASSED THROUGH THE URETHRA AND ENTERED THE FORCED OF TWO CENTIMETERS INTO THE URETER, AND WHEN THE STENT WAS PUSHED, IT BECAME BUCKLED AND WOULD NOT ADVANCE INTO THE URETER AND A TORN WAS OBSERVED ON THE DISTAL PART. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER CONTOUR VL URETERAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14950119 ↗
  • MalfunctionFeb 18, 2022Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CONTOUR VL URETERAL STENT WAS USED DURING A STENT PLACEMENT PROCEDURE IN THE URETER. THE EVENT DATE WAS NOT REPORTED. DURING PREPARATION, IT WAS NOTED THAT THE CONTOUR VL URETERAL STENT WAS FOUND SEPARATED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER CONTOUR VL URETERAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE AS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13559738 ↗

Other STENT, URETERAL devices

All STENT, URETERAL devices →