Percuflex™ Plus M0061752540Recalls & reported problems

Boston Scientific Corporation · Ureteral Stent · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Boston Scientific Percuflex Plus ureteral stent is a small tube placed in the ureter to keep it open. The 35 reports mostly describe the stent becoming encrusted with mineral deposits so that it had to be removed, with one report in 2024 noting a single stent found encrusted and one earlier report describing about 30 encrusted stents found in a patient within two weeks. Reports mention 30 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
35
Deaths reported
0
Injuries reported
30
Recalls naming it
0

Summary

The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 08714729067757), received between Dec 5, 2016 and Jul 18, 2024 — classified as 30 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Calcified” (33 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Calcified33
  • Break2
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions33
  • No Consequences Or Impact To Patient2
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 18, 2024Calcified; Calcified

    IT WAS REPORTED THAT, DURING A TRANSURETHRAL LITHOTOMY PROCEDURE, A PERCUFLEX PLUS URETERAL STENT WAS SUCCESSFULLY IMPLANTED ON (B)(6) 2024. ON (B)(6) 2024, IT WAS FOUND THAT THE STENT WAS ENCRUSTED, AND IT WAS REMOVED FROM THE PATIENT. ANOTHER OF THE SAME STENT WAS USED TO REPLACE THE ENCRUSTED STENT. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19779685 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETHRAL LITHOTOMY PROCEDURE IN THE URETER. THE PROCEDURE DATE IS UNKNOWN. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON JANUARY 23, 2023, STATING THE IMPLANT DATES FOR THE AFFECTED STENTS OCCURRED OVER SEVERAL MONTHS AND NOT ALL WERE PERFORMED BY THE SAME PHYSICIAN. IT WAS ALSO REPORTED THAT SOME OF THE STENTS WERE STILL ABLE TO DRAIN EVEN THOUGH THEY WERE FOUND ENCRUSTED. THE PATIENTS UNDERWENT REGULAR CHECKUPS AND AN UNSPECIFIED NUMBER OF PATIENTS HAD MEDICAL CONDITIONS OR MEDICATION THAT COULD HAVE CONTRIBUTED TO THE ENCRUSTATION. SOME OF THE PATIENTS UNDERWENT DIAGNOSTIC TESTING TO CONFIRM THEY HAD NO METABOLIC OR ELECTROLYTE PROBLEMS. THE STENTS WERE REMOVED EARLIER THAN THE SCHEDULED DATES WITH THE USE OF PNEUMATIC SYSTEM TO BREAK THE STONES AND ENCRUSTATION. SOME OF THE STENTS WERE REPORTED TO HAVE BEEN TORN, BUT WITH THE PNEUMATIC SYSTEM ALL PIECES OF THE STENTS WERE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16164894 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETHRAL LITHOTOMY PROCEDURE IN THE URETER. THE PROCEDURE DATE IS UNKNOWN. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON JANUARY 23, 2023, STATING THE IMPLANT DATES FOR THE AFFECTED STENTS OCCURRED OVER SEVERAL MONTHS AND NOT ALL WERE PERFORMED BY THE SAME PHYSICIAN. IT WAS ALSO REPORTED THAT SOME OF THE STENTS WERE STILL ABLE TO DRAIN EVEN THOUGH THEY WERE FOUND ENCRUSTED. THE PATIENTS UNDERWENT REGULAR CHECKUPS AND AN UNSPECIFIED NUMBER OF PATIENTS HAD MEDICAL CONDITIONS OR MEDICATION THAT COULD HAVE CONTRIBUTED TO THE ENCRUSTATION. SOME OF THE PATIENTS UNDERWENT DIAGNOSTIC TESTING TO CONFIRM THEY HAD NO METABOLIC OR ELECTROLYTE PROBLEMS. THE STENTS WERE REMOVED EARLIER THAN THE SCHEDULED DATES WITH THE USE OF PNEUMATIC SYSTEM TO BREAK THE STONES AND ENCRUSTATION. SOME OF THE STENTS WERE REPORTED TO HAVE BEEN TORN, BUT WITH THE PNEUMATIC SYSTEM ALL PIECES OF THE STENTS WERE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16164803 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETERAL LIHTOTOMY PROCEDURE IN THE URETER. THE EVENT DATE WAS NOT REPORTED. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16164684 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETERAL LIHTOTOMY PROCEDURE IN THE URETER. THE EVENT DATE WAS NOT REPORTED. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16164207 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETERAL LIHTOTOMY PROCEDURE IN THE URETER. THE EVENT DATE WAS NOT REPORTED. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON JANUARY 23, 2023, STATING THE IMPLANT DATES FOR THE AFFECTED STENTS OCCURRED OVER SEVERAL MONTHS AND NOT ALL WERE PERFORMED BY THE SAME PHYSICIAN. IT WAS ALSO REPORTED THAT SOME OF THE STENTS WERE STILL ABLE TO DRAIN EVEN THOUGH THEY WERE FOUND ENCRUSTED. THE PATIENTS UNDERWENT REGULAR CHECKUPS AND AN UNSPECIFIED NUMBER OF PATIENTS HAD MEDICAL CONDITIONS OR MEDICATION THAT COULD HAVE CONTRIBUTED TO THE ENCRUSTATION. SOME OF THE PATIENTS UNDERWENT DIAGNOSTIC TESTING TO CONFIRM THEY HAD NO METABOLIC OR ELECTROLYTE PROBLEMS. THE STENTS WERE REMOVED EARLIER THAN THE SCHEDULED DATES WITH THE USE OF PNEUMATIC SYSTEM TO BREAK THE STONES AND ENCRUSTATION. SOME OF THE STENTS WERE REPORTED TO HAVE BEEN TORN, BUT WITH THE PNEUMATIC SYSTEM ALL PIECES OF THE STENTS WERE REMOVE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16164111 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETHRAL LITHOTOMY PROCEDURE IN THE URETER. THE PROCEDURE DATE IS UNKNOWN. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON JANUARY 23, 2023, STATING THE IMPLANT DATES FOR THE AFFECTED STENTS OCCURRED OVER SEVERAL MONTHS AND NOT ALL WERE PERFORMED BY THE SAME PHYSICIAN. IT WAS ALSO REPORTED THAT SOME OF THE STENTS WERE STILL ABLE TO DRAIN EVEN THOUGH THEY WERE FOUND ENCRUSTED. THE PATIENTS UNDERWENT REGULAR CHECKUPS AND AN UNSPECIFIED NUMBER OF PATIENTS HAD MEDICAL CONDITIONS OR MEDICATION THAT COULD HAVE CONTRIBUTED TO THE ENCRUSTATION. SOME OF THE PATIENTS UNDERWENT DIAGNOSTIC TESTING TO CONFIRM THEY HAD NO METABOLIC OR ELECTROLYTE PROBLEMS. THE STENTS WERE REMOVED EARLIER THAN THE SCHEDULED DATES WITH THE USE OF PNEUMATIC SYSTEM TO BREAK THE STONES AND ENCRUSTATION. SOME OF THE STENTS WERE REPORTED TO HAVE BEEN TORN, BUT WITH THE PNEUMATIC SYSTEM ALL PIECES OF THE STENTS WERE REMOVED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16163844 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETERAL LIHTOTOMY PROCEDURE IN THE URETER. THE EVENT DATE WAS NOT REPORTED. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16162435 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETHERAL LITHOTOMY PROCEDURE IN THE URETER. THE EVENT DATE WAS NOT REPORTED. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16162406 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETERAL LITHOTOMY PROCEDURE IN THE URETER. THE EVENT DATE WAS NOT REPORTED. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON JANUARY 23, 2023, STATING THE IMPLANT DATES FOR THE AFFECTED STENTS OCCURRED OVER SEVERAL MONTHS AND NOT ALL WERE PERFORMED BY THE SAME PHYSICIAN. IT WAS ALSO REPORTED THAT SOME OF THE STENTS WERE STILL ABLE TO DRAIN EVEN THOUGH THEY WERE FOUND ENCRUSTED. THE PATIENTS UNDERWENT REGULAR CHECKUPS AND AN UNSPECIFIED NUMBER OF PATIENTS HAD MEDICAL CONDITIONS OR MEDICATION THAT COULD HAVE CONTRIBUTED TO THE ENCRUSTATION. SOME OF THE PATIENTS UNDERWENT DIAGNOSTIC TESTING TO CONFIRM THEY HAD NO METABOLIC OR ELECTROLYTE PROBLEMS. THE STENTS WERE REMOVED EARLIER THAN THE SCHEDULED DATES WITH THE USE OF PNEUMATIC SYSTEM TO BREAK THE STONES AND ENCRUSTATION. SOME OF THE STENTS WERE REPORTED TO HAVE BEEN TORN, BUT WITH THE PNEUMATIC SYSTEM ALL PIECES OF THE STENTS WERE REMOVE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16162329 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETERAL LIHTOTOMY PROCEDURE IN THE URETER. THE EVENT DATE WAS NOT REPORTED. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16162320 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETERAL LIHTOTOMY PROCEDURE IN THE URETER. THE EVENT DATE WAS NOT REPORTED. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16162310 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETERAL LIHTOTOMY PROCEDURE IN THE URETER. THE EVENT DATE WAS NOT REPORTED. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16162301 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETHERAL LITHOTOMY PROCEDURE IN THE URETER. THE EVENT DATE WAS NOT REPORTED. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16162298 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETERAL LIHTOTOMY PROCEDURE IN THE URETER. THE EVENT DATE WAS NOT REPORTED. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16162251 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETERAL LITHOTOMY PROCEDURE IN THE URETER. THE EVENT DATE WAS NOT REPORTED. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON JANUARY 23, 2023, STATING THE IMPLANT DATES FOR THE AFFECTED STENTS OCCURRED OVER SEVERAL MONTHS AND NOT ALL WERE PERFORMED BY THE SAME PHYSICIAN. IT WAS ALSO REPORTED THAT SOME OF THE STENTS WERE STILL ABLE TO DRAIN EVEN THOUGH THEY WERE FOUND ENCRUSTED. THE PATIENTS UNDERWENT REGULAR CHECKUPS AND AN UNSPECIFIED NUMBER OF PATIENTS HAD MEDICAL CONDITIONS OR MEDICATION THAT COULD HAVE CONTRIBUTED TO THE ENCRUSTATION. SOME OF THE PATIENTS UNDERWENT DIAGNOSTIC TESTING TO CONFIRM THEY HAD NO METABOLIC OR ELECTROLYTE PROBLEMS. THE STENTS WERE REMOVED EARLIER THAN THE SCHEDULED DATES WITH THE USE OF PNEUMATIC SYSTEM TO BREAK THE STONES AND ENCRUSTATION. SOME OF THE STENTS WERE REPORTED TO HAVE BEEN TORN, BUT WITH THE PNEUMATIC SYSTEM ALL PIECES OF THE STENTS WERE REMOVE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16162223 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETERAL LIHTOTOMY PROCEDURE IN THE URETER. THE EVENT DATE WAS NOT REPORTED. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON JANUARY 23, 2023, STATING THE IMPLANT DATES FOR THE AFFECTED STENTS OCCURRED OVER SEVERAL MONTHS AND NOT ALL WERE PERFORMED BY THE SAME PHYSICIAN. IT WAS ALSO REPORTED THAT SOME OF THE STENTS WERE STILL ABLE TO DRAIN EVEN THOUGH THEY WERE FOUND ENCRUSTED. THE PATIENTS UNDERWENT REGULAR CHECKUPS AND AN UNSPECIFIED NUMBER OF PATIENTS HAD MEDICAL CONDITIONS OR MEDICATION THAT COULD HAVE CONTRIBUTED TO THE ENCRUSTATION. SOME OF THE PATIENTS UNDERWENT DIAGNOSTIC TESTING TO CONFIRM THEY HAD NO METABOLIC OR ELECTROLYTE PROBLEMS. THE STENTS WERE REMOVED EARLIER THAN THE SCHEDULED DATES WITH THE USE OF PNEUMATIC SYSTEM TO BREAK THE STONES AND ENCRUSTATION. SOME OF THE STENTS WERE REPORTED TO HAVE BEEN TORN, BUT WITH THE PNEUMATIC SYSTEM ALL PIECES OF THE STENTS WERE REMOVE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16162211 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETHRAL LITHOTOMY PROCEDURE IN THE URETER. THE PROCEDURE DATE IS UNKNOWN. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON JANUARY 23, 2023, STATING THE IMPLANT DATES FOR THE AFFECTED STENTS OCCURRED OVER SEVERAL MONTHS AND NOT ALL WERE PERFORMED BY THE SAME PHYSICIAN. IT WAS ALSO REPORTED THAT SOME OF THE STENTS WERE STILL ABLE TO DRAIN EVEN THOUGH THEY WERE FOUND ENCRUSTED. THE PATIENTS UNDERWENT REGULAR CHECKUPS AND AN UNSPECIFIED NUMBER OF PATIENTS HAD MEDICAL CONDITIONS OR MEDICATION THAT COULD HAVE CONTRIBUTED TO THE ENCRUSTATION. SOME OF THE PATIENTS UNDERWENT DIAGNOSTIC TESTING TO CONFIRM THEY HAD NO METABOLIC OR ELECTROLYTE PROBLEMS. THE STENTS WERE REMOVED EARLIER THAN THE SCHEDULED DATES WITH THE USE OF PNEUMATIC SYSTEM TO BREAK THE STONES AND ENCRUSTATION. SOME OF THE STENTS WERE REPORTED TO HAVE BEEN TORN, BUT WITH THE PNEUMATIC SYSTEM ALL PIECES OF THE STENTS WERE REMOVED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16161569 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETHRAL LITHOTOMY PROCEDURE IN THE URETER. THE PROCEDURE DATE IS UNKNOWN. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON JANUARY 23, 2023, STATING THE IMPLANT DATES FOR THE AFFECTED STENTS OCCURRED OVER SEVERAL MONTHS AND NOT ALL WERE PERFORMED BY THE SAME PHYSICIAN. IT WAS ALSO REPORTED THAT SOME OF THE STENTS WERE STILL ABLE TO DRAIN EVEN THOUGH THEY WERE FOUND ENCRUSTED. THE PATIENTS UNDERWENT REGULAR CHECKUPS AND AN UNSPECIFIED NUMBER OF PATIENTS HAD MEDICAL CONDITIONS OR MEDICATION THAT COULD HAVE CONTRIBUTED TO THE ENCRUSTATION. SOME OF THE PATIENTS UNDERWENT DIAGNOSTIC TESTING TO CONFIRM THEY HAD NO METABOLIC OR ELECTROLYTE PROBLEMS. THE STENTS WERE REMOVED EARLIER THAN THE SCHEDULED DATES WITH THE USE OF PNEUMATIC SYSTEM TO BREAK THE STONES AND ENCRUSTATION. SOME OF THE STENTS WERE REPORTED TO HAVE BEEN TORN, BUT WITH THE PNEUMATIC SYSTEM ALL PIECES OF THE STENTS WERE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16161444 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETHRAL LITHOTOMY PROCEDURE IN THE URETER. THE PROCEDURE DATE IS UNKNOWN. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16161421 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETHRAL LITHOTOMY PROCEDURE IN THE URETER. THE PROCEDURE DATE IS UNKNOWN. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON JANUARY 23, 2023, STATING THE IMPLANT DATES FOR THE AFFECTED STENTS OCCURRED OVER SEVERAL MONTHS AND NOT ALL WERE PERFORMED BY THE SAME PHYSICIAN. IT WAS ALSO REPORTED THAT SOME OF THE STENTS WERE STILL ABLE TO DRAIN EVEN THOUGH THEY WERE FOUND ENCRUSTED. THE PATIENTS UNDERWENT REGULAR CHECKUPS AND AN UNSPECIFIED NUMBER OF PATIENTS HAD MEDICAL CONDITIONS OR MEDICATION THAT COULD HAVE CONTRIBUTED TO THE ENCRUSTATION. SOME OF THE PATIENTS UNDERWENT DIAGNOSTIC TESTING TO CONFIRM THEY HAD NO METABOLIC OR ELECTROLYTE PROBLEMS. THE STENTS WERE REMOVED EARLIER THAN THE SCHEDULED DATES WITH THE USE OF PNEUMATIC SYSTEM TO BREAK THE STONES AND ENCRUSTATION. SOME OF THE STENTS WERE REPORTED TO HAVE BEEN TORN, BUT WITH THE PNEUMATIC SYSTEM ALL PIECES OF THE STENTS WERE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16161282 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETHRAL LITHOTOMY PROCEDURE IN THE URETER. THE PROCEDURE DATE IS UNKNOWN. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON JANUARY 23, 2023, STATING THE IMPLANT DATES FOR THE AFFECTED STENTS OCCURRED OVER SEVERAL MONTHS AND NOT ALL WERE PERFORMED BY THE SAME PHYSICIAN. IT WAS ALSO REPORTED THAT SOME OF THE STENTS WERE STILL ABLE TO DRAIN EVEN THOUGH THEY WERE FOUND ENCRUSTED. THE PATIENTS UNDERWENT REGULAR CHECKUPS AND AN UNSPECIFIED NUMBER OF PATIENTS HAD MEDICAL CONDITIONS OR MEDICATION THAT COULD HAVE CONTRIBUTED TO THE ENCRUSTATION. SOME OF THE PATIENTS UNDERWENT DIAGNOSTIC TESTING TO CONFIRM THEY HAD NO METABOLIC OR ELECTROLYTE PROBLEMS. THE STENTS WERE REMOVED EARLIER THAN THE SCHEDULED DATES WITH THE USE OF PNEUMATIC SYSTEM TO BREAK THE STONES AND ENCRUSTATION. SOME OF THE STENTS WERE REPORTED TO HAVE BEEN TORN, BUT WITH THE PNEUMATIC SYSTEM ALL PIECES OF THE STENTS WERE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16161204 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETHRAL LITHOTOMY PROCEDURE IN THE URETER, PERFORMED ON (B)(6) 2022. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON JANUARY 23, 2023, STATING THE IMPLANT DATES FOR THE AFFECTED STENTS OCCURRED OVER SEVERAL MONTHS AND NOT ALL WERE PERFORMED BY THE SAME PHYSICIAN. IT WAS ALSO REPORTED THAT SOME OF THE STENTS WERE STILL ABLE TO DRAIN EVEN THOUGH THEY WERE FOUND ENCRUSTED. THE PATIENTS UNDERWENT REGULAR CHECKUPS AND AN UNSPECIFIED NUMBER OF PATIENTS HAD MEDICAL CONDITIONS OR MEDICATION THAT COULD HAVE CONTRIBUTED TO THE ENCRUSTATION. SOME OF THE PATIENTS UNDERWENT DIAGNOSTIC TESTING TO CONFIRM THEY HAD NO METABOLIC OR ELECTROLYTE PROBLEMS. THE STENTS WERE REMOVED EARLIER THAN THE SCHEDULED DATES WITH THE USE OF PNEUMATIC SYSTEM TO BREAK THE STONES AND ENCRUSTATION. SOME OF THE STENTS WERE REPORTED TO HAVE BEEN TORN, BUT WITH THE PNEUMATIC SYSTEM ALL PIECES OF THE STENTS WERE REMOVED. IT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16161158 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETHRAL LITHOTOMY PROCEDURE IN THE URETER. THE PROCEDURE DATE IS UNKNOWN. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16161069 ↗
  • InjuryJan 13, 2023Calcified; Calcified

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUFLEX PLUS URETERAL STENT WAS USED DURING A TRANSURETHRAL LITHOTOMY PROCEDURE IN THE URETER. THE PROCEDURE DATE IS UNKNOWN. POST PROCEDURE, IT WAS REPORTED THAT THERE WERE APPROXIMATELY 30 STENTS THAT WERE FOUND ENCRUSTED IN THE PATIENT'S BODY IN TWO WEEKS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16161022 ↗