Skip to content

Percuflex™ Nephroureteral Stent M001221370Recalls & reported problems

Boston Scientific Corporation · Nephroureteral Stent System · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729114086), received between May 16, 2023 and Aug 22, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 22, 2024Break; Break

    IT WAS REPORTED THAT CATHETER BREAK OCCURRED. A PERCUFLEX NEPHROURETERAL STENT WAS SELECTED FOR USE FOR A LEFT ANTEGRADE URETERIC STENTING. DURING THE PROCEDURE, A GUIDEWIRE WAS INSERTED AND PLACED INTO THE URINARY SYSTEM. CONSEQUENTLY, A PERCUFLEX WAS ADVANCED. HOWEVER, WHEN STRING WAS PULLED, THE CATHETER SPLIT INTO TWO. THE DEVICE WAS FULLY REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20044181 ↗
  • MalfunctionMay 16, 2023Break; Break

    IT WAS REPORTED THAT THE STENT DETACHMENT OCCURRED. A PERCUFLEX NEPHROURETERAL STENT WAS SELECTED FOR USE IN THE URINARY BLADDER. DURING THE PROCEDURE, THE NEPHROURETERAL STENT WAS ADVANCED OVER THE GUIDEWIRE INTO THE URINARY BLADDER. THE GUIDEWIRE WAS REMOVED TO FORM THE NEPHROURETERAL STENT. THE SUTURE WAS PULLED TO CREATE THE PIGTAIL AND UPON REMOVING OVER THE WIRE, IT WAS NOTED THAT A FRACTURE OCCURRED AT THE PROXIMAL END OF THE CATHETER DISTAL TO THE HUB AND AT THE DISTAL THIRD. THE GUIDE WIRE WAS READVANCED TO REMOVE THE NEPHROURETERAL STENT AND RESISTANCE WAS MET. THE NEPHROURETERAL STENT WAS REMOVED OVER THE WIRE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16938428 ↗

Other STENT, URETERAL devices

All STENT, URETERAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page