Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of Sterling (153)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729122937), received between Oct 2, 2019 and Jul 3, 2024 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture5
  • Break1
  • Difficult to Remove1
  • Leak/Splash1
  • Unsealed Device Packaging1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 3, 2024Leak/Splash; Leak/Splash

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 12JUN2024. IT WAS REPORTED THAT BALLOON LEAK OCCURRED. THE TARGET LOCATION WAS LOCATED IN THE LOWER EXTREMITY ARTERY. A 5.0MMX60MMX135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, IT WAS NOTED THAT THE TIP OF THE BALLOON WAS LEAKING. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS STABLE. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED A BALLOON TEAR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19669074 ↗
  • MalfunctionApr 25, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND SEVERELY CALCIFIED PERIPHERAL ARTERY. A 5.0MMX60MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED. THE DEVICE WAS TAKEN OUT AND DISPOSED AND REPLACED FOR A DIFFERENT DEVICE TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED. IT WAS FURTHER REPORTED THAT THE BALLOON RUPTURED ON THE FIRST INFLATION WHEN IT WAS IN THE PROCESS OF INFLATING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19181489 ↗
  • MalfunctionSep 7, 2023Break; Difficult to Remove; Material Rupture; Break; Difficult to Remove; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE, SHAFT BREAK AND REMOVAL DIFFICULTIES OCCURRED. THE PATIENT PRESENTED WITH PERIPHERAL ARTERY DISEASE (PAD) AND ANGIOGRAM WAS PERFORMED TO FIX A HEAVILY CALCIFIED CHRONIC TOTAL OCCLUSION (CTO) IN EXTERNAL ILIAC TO SUPERFICIAL FEMORAL ARTERY THE 100% STENOSED TARGET LESION WAS MODERATELY TORTUOUS AND SEVERELY CALCIFIED. A 5.0MMX60MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INITIAL INFLATION AT 14 ATMOSPHERES FOR 10 SECONDS, THE BALLOON BURST. THE BALLOON WAS THEN ATTEMPTED TO BE REMOVED BUT IT WAS DIFFICULT; HENCE, THE PHYSICIAN PULLED THE DEVICE WITH FORCE AND THE BALLOON SNAPPED IN HALF. SINCE THE BALLOON HAD BURST; DOUBLE NEGATIVE PRESSURE WAS APPLIED TO DEFLATE AND PULLED THE BALLOON IN AS CLOSE AS POSSIBLE TO THE SHAFT. TOGETHER WITH THE WIRE, THE BALLOON WAS FRACTURED WHEN REMOVED SUCCESSFULLY AND NO DEVICE FRAGMENTS LEFT INSIDE THE PATIENT. A NEW STERLING BALLOON CATHETER WAS USED, AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17698100 ↗
  • MalfunctionFeb 13, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT WAS BEING TREATED FOR ANGIOGRAM IN THE LEG. A 5.0MMX60MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, WHILE INFLATING THE BALLOON AT NOMINAL PRESSURE DURING PROCEDURE, THE BALLOON LEAKED AND POPPED. THE DEVICE WAS REMOVED FROM THE BODY. A NEW OF SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16362578 ↗
  • MalfunctionDec 5, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 97% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 5.0MMX60MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, ON INITIAL INFLATION AT 10 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER 5.0 X 60 STERLING BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12931339 ↗
  • MalfunctionDec 9, 2020Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 5.0MMX60MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION AT NOMINAL PRESSURE, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10976778 ↗
  • MalfunctionOct 2, 2019Unsealed Device Packaging; Unsealed Device Packaging

    IT WAS REPORTED THAT STERILITY WAS COMPROMISED. DURING PREPARATION OF A 5.0MM X 60MM X 135CM (4F) STERLING BALLOON CATHETER, IT WAS NOTICED THAT THE INNER DEVICE POUCH WAS SLIGHTLY OPEN UPON INSPECTION. THE DEVICE WAS NOT USED INSIDE THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED.

    Patient problems: No Patient Involvement

    MAUDE report 9146380 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page