Sterling™ H74939032104080Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of Sterling (153)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 08714729123453), received between Jan 28, 2020 and Oct 22, 2025 — classified as 1 injury, 14 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture14
- Detachment of Device or Device Component1
- Difficult to Remove1
- Inflation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- No Consequences Or Impact To Patient3
- Foreign Body In Patient1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 22, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED RIGHT SUPERFICIAL FEMORAL ARTERY. A 10.0MMX40MMX80CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER THE SECOND INFLATION DURING PRESSURIZATION. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23358699 ↗MalfunctionAug 22, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS AN AREA OF IN-STENT RESTENOSIS (ISR) LOCATED IN THE MODERATELY TORTUOUS ARTERIO-VENOUS FISTULA. A 10.0MMX40MMX80CM (5F) STERLING BALLOON CATHETER WAS USED TO DILATE A PREVIOUSLY IMPLANTED EPIC STENT. HOWEVER, DURING THE 2ND INFLATION AT 10 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22877843 ↗MalfunctionFeb 11, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT COMMON ILIAC ARTERY. A 10.0MMX40MMX80CM (5F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE SECOND INFLATION AT 10 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21358507 ↗InjurySep 3, 2024Material Rupture; Detachment of Device or Device Component; Material Rupture; Detachment of Device or Device Component
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
Patient problems: Foreign Body In Patient
MAUDE report 20128973 ↗MalfunctionJul 24, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 10.0MMX40MMX80CM (5F) STERLING BALLOON CATHETER WAS ADVANCED FOR A SHUNT PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY. DURING THE PROCEDURE, ON THE FIRST INFLATION AT 8 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM AND NO DAMAGE. A DIFFERENT MODEL WAS USED TO COMPLETE THE PROCEDURE. NO COMPLICATIONS REPORTED, AND PATIENT STATUS WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19817269 ↗MalfunctionMay 28, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT COMMON ILIAC ARTERY. A 10.0MMX40MMX80CM (5F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE SECOND INFLATION AT 10 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS IN GOOD CONDITION POST PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17016117 ↗MalfunctionMay 23, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED COMMON ILIAC ARTERY. A 10.0MMX40MMX80CM (5F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE SECOND INFLATION AT 6 ATMOSPHERES FOR 5 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND PATIENT CONDITION WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16985434 ↗MalfunctionMay 5, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED COMMON ILIAC ARTERY. A 10.0MMX40MMX80CM (5F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED AT THE MIDDLE AND A PINHOLE WAS NOTED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16878373 ↗MalfunctionFeb 20, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED COMMON ILIAC ARTERY. A 10.0MM X 40MM X 80CM (5F) STERLING OTW BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 15 SECONDS, THE BALLOON RUPTURED VERTICALLY AT THE CENTER. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT CONDITION WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16403020 ↗MalfunctionApr 28, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 75% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND SEVERELY CALCIFIED RIGHT COMMON ILIAC ARTERY. AFTER INSERTING A NON-BOSTON SCIENTIFIC (BSC) INTRODUCER SHEATH, THE LESION WAS CROSSED WITH A JUPITER FC GUIDEWIRE. A 10.0MMX40MMX80CM (5F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A NON-BSC DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS IN GOOD CONDITION POST PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14245147 ↗MalfunctionAug 5, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN THE MODERATELY TORTUOUS SUBCLAVIAN ARTERY. A 10.0MMX40MMX80CM (5F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT INJURIES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12284967 ↗MalfunctionMay 3, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND NON-CALCIFIED SUBCLAVIAN VEIN. A 10.0MMX40MMX80CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE INITIAL INFLATION AT 6 ATMOSPHERES FOR 6 SECONDS, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11766654 ↗MalfunctionJul 8, 2020Difficult to Remove; Material Rupture; Difficult to Remove; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED AND REMOVAL DIFFICULTIES WERE ENCOUNTERED. THE TARGET LESION WAS LOCATED IN THE SUBCLAVIAN VEIN. A 10.0MM X 40MM X 80CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED HORIZONTALLY. RESISTANCE WAS ENCOUNTERED UPON REMOVING THE DEVICE, SO THE DEVICE WAS REMOVED TOGETHER WITH THE SHEATH. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S STATUS WAS GOOD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10244300 ↗MalfunctionMar 12, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MILDLY CALCIFIED SUBCLAVIAN VEIN. AFTER A GUIDEWIRE CROSSED THE LESION, A 10.0MMX40MMX80CM STERLING BALLOON CATHETER WAS ADVANCED FOR PRE-DILATATION. HOWEVER, DURING THE SECOND INFLATION, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9823123 ↗MalfunctionJan 28, 2020Inflation Problem; Inflation Problem
REPORTABLE BASED ON THE DEVICE ANALYSIS ON 09JAN2020. IT WAS REPORTED THAT INFLATION FAILURE OCCURRED. VASCULAR ACCESS WAS OBTAINED VIA BRACHIAL APPROACH. THE 75% STENOSED WITH LARGE VESSEL DIAMETER TARGET LESION WAS LOCATED IN THE LEFT SUBCLAVIAN VESSEL. A 10.0MMX40MMX80CM (5F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, THE BALLOON WAS UNABLE TO INFLATE. THE PHYSICIAN ATTEMPTED TO DILATE THE LESION BY DOUBLE BALLOON USING TWO 7.0 X 40, 75CM MUSTANG BALLOON CATHETERS. HOWEVER, DURING THE FIRST INFLATION AT 18 ATMOSPHERES, ONE OF THE BALLOONS RUPTURED. BOTH BALLOONS WERE REMOVED AND THE PROCEDURE WAS COMPLETED WITH 10MM STERLING AND 7MM MUSTANG BALLOON CATHETERS. THERE WERE NO PATIENT COMPLICATIONS REPORTED. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED A LONGITUDINAL TEAR ON THE BALLOON.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9638324 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page