Skip to content

Sterling™ H74939032406010Recalls & reported problems

Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of Sterling (153)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729123644), received between Apr 9, 2020 and Jan 14, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture3
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Deflation Problem1
  • Inflation Problem1
  • Unsealed Device Packaging1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Patient Involvement1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 14, 2026Inflation Problem; Adverse Event Without Identified Device or Use Problem

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 17DEC2025. IT WAS REPORTED THAT BALLOON DID NOT FULLY EXPAND. THE SEVERELY STENOSED TARGET LESION WAS LOCATED IN THE ARTERIOVENOUS FISTULA. A 4.0MMX60MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON WAS INFLATED UP TO 23 ATMOSPHERES, HOWEVER THE BALLOON DID NOT FULLY EXPAND IN THE LESION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. HOWEVER, DEVICE ANALYSIS REVEALED THAT THE BALLOON IS SEPARATED 7MM FROM THE TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24069315 ↗
  • MalfunctionMay 25, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 4.0MMX60MMX135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING FIRST INFLATION, THE BALLOON BEGAN LEAKING CONTRAST AND NOTED TO BE RUPTURED. THE PROCEDURE WAS COMPLETED WITH A 4X220 STERLING BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17007089 ↗
  • MalfunctionApr 21, 2023Break; Break

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. DURING UNPACKING OF A 4.0MMX60MMX135CM (4F) FG STERLING BALLOON CATHETER, IT WAS FOUND THAT THE BALLOON WAS BROKEN. THE DEVICE WAS NEVER USED INSIDE THE PATIENT'S BODY. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16782701 ↗
  • MalfunctionOct 20, 2022Deflation Problem; Deflation Problem

    IT WAS REPORTED THAT THE BALLOON HAD DEFLATION ISSUES. A STERLING BALLOON CATHETER 4.00MM X 60MM, 135CM WAS SELECTED FOR USE TO TREAT PERIPHERAL ARTERIAL OCCLUSIVE DISEASE BY PERFORMING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE. THE PHYSICIAN INFLATED THE BALLOON TO NOMINAL PRESSURE; HOWEVER, THE BALLOON WAS NOT ABLE TO BE DEFLATED. THE CATHETER WAS ABLE TO BE REMOVED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING ANOTHER STERLING DEVICE WITHOUT SEQUALA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15644394 ↗
  • MalfunctionApr 7, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE RIGHT LEG. A 4.0MMX60MMX135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14044747 ↗
  • MalfunctionMar 8, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE HEAVILY DISEASED TARGET LESION WAS LOCATED IN THE DISTAL SUPERFICIAL FEMORAL ARTERY. A 4.0MMX60MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION POST-LASER. HOWEVER, DURING THE PROCEDURE, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER SMALLER BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13691145 ↗
  • MalfunctionApr 9, 2020Unsealed Device Packaging; Unsealed Device Packaging

    IT WAS REPORTED THAT THERE WAS A DEVICE PACKAGING ISSUE. A 4.0MMX60MMX135CM (4F) STERLING BALLOON CATHETER WAS SELECTED FOR USE. IT APPEARED THAT THE DEVICE LOOKED LIKE IT WAS ALREADY OPENED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Patient Involvement

    MAUDE report 9943753 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page